TROUBLESHOOTING GUIDE
FOR PR-23 ALL MODELS

10 (14)

January 9, 2015

K-PATENTS OY

Postal Address: P.O.Box 77, FI-01511 Vantaa, Finland

Street Address: Elannontie 5, FI-01510 Vantaa, Finland

Tel. int.+358 207 291 570

Fax int.+358 207 291 577

[email protected]

www.kpatents.com

Vat No. FI03035575

Business ID 0303557-5

Registered in Helsinki

increase. The visible changes in the curve and the temperature depend on viscosity, steam pressure and

temperature, as well as the

software version of the transmitter. If the wash is not reaching the prism, whole

wash system should be examined. There might be blockage in the lines, the wash pressure might be too low

or there might be condensate in the lines because lack of insulation.

Figure 9

Example of successful prism wash

Best way to diagnose and verify successful washing is to log data from the instrument with FC-11. The FC-

11 logs all the diagnostic values, as well as, all the measurement values. As it can be seen from the picture,

the temperature (red) increases and the CALC (blue) decreases with every wash, whereas the CONC

(green) stays stable all the time. This also reveals intermittent problems in the washing systems.

Figure 10

Data shows clearly if the wash is successful or not.