Specifications, Analog input, Chapter 5 – Measurement Computing USB-2416-4AO User Manual

Page 22

Advertising
background image

22

Chapter 5

Specifications

All specifications are subject to change without notice.
Typical for 25 °C unless otherwise specified.
All specifications apply to all temperature and voltage input channels unless otherwise specified.
Specifications in italic
text are guaranteed by design.

Analog input

Table 1. General analog input specifications

Parameter

Conditions

Specification

A/D converter type

ADS1256, 24-bit Sigma Delta

A/D data rates

3750 samples per second (S/s), 2000 S/s, 1000 S/s, 500 S/s, 100 S/s,
60 S/s, 50 S/s, 25 S/s, 10 S/s, 5 S/s, 2.5 S/s

Throughput

Single channel: 2.5 Hz to 1102.94 Hz, software selectable

Multiple channels: 0.16 Hz to 1102.94 Hz, software selectable

See Table 18 and Table 19 for details.

Number of channels

Up to 32 channels individually software-configurable as single-
ended or differential
Thermocouples require differential mode. For each channel
configured as differential, you essentially lose one single-ended
channel. You can add channels by connecting to an AI-EXP32.

Input isolation

500 VDC minimum between field wiring and USB interface

Channel configurations

Temperature sensor input, software programmable to match sensor
type

Voltage input

Input voltage range

Thermocouple mode

±0.078125 V

Voltage mode (Note 1)

±20 V, ±10 V, ±5 V, ±2.5 V, ±1.25 V, ±0.625 V, ±0.3125 V,
±0.15625 V, ±0.078125 V, software-configurable

Absolute maximum
input voltage

CxH-CxL relative to
GND

±30 V maximum (power on)

±10 V maximum (power off)

Input impedance

2 G

Ω (power on)

390Ω (power off)

Input leakage current

±10.6 nA

Input capacitance

590 pf

Maximum working
voltage (signal +
common mode)

Voltage mode: ±20 V
range

±20.01 V maximum

Voltage mode: all other
voltage input ranges

±10.25 V maximum

Common mode
rejection ratio (Note 1)

Thermocouple mode,
(f

IN

= 60 Hz)

110 dB

Voltage mode,(f

IN

=

60 Hz, all input ranges)

90 dB

ADC resolution

24 bits

Crosstalk

Adjacent channels

100 dB

Input coupling

DC

Channel gain queue

Up to 64 elements

Software-configurable channel and range

Warm-up time

45 minutes minimum

Open thermocouple
detect

Automatically enabled when the channel is configured for a
thermocouple sensor.

CJC sensor accuracy

15 °C to 35 °C

±0.15 °C typical

0 °C to 55 °C

±0.5 °C maximum

Advertising