Measurement Computing DBK52 User Manual

Dbk52, Channel thermocouple input module, Overview

Advertising
background image

DBK52

14-Channel Thermocouple Input Module

Overview …… 1
Hardware Setup …… 2

Module Connection …… 2
Module Configuration …… 3
DaqBook/100 Series & /200 Series and DaqBoard [ISA type] Configuration …… 3
DaqBook/2000 Series and DaqBoard/2000 Series Configuration …… 3

Software Setup …… 4
DBK52 – Specifications …… 4

Reference Notes:

o

Chapter 2 includes pinouts for P1, P2, P3, and P4. Refer to pinouts applicable to

your system, as needed.

o

In regard to calculating system power requirements, refer to DBK Basics located

near the front of this manual.

Overview

DaqView Users: When DBK52 is used with /2000 Series Devices, the Internal Clock Speed
should be set to 100 kHz as described in Chapter 3, DBK Setup in DaqView
.

The DBK52 provides the ability to scan and input thermocouple data for various computer-based
temperature-monitoring systems. Features include:

• Each DBK52 can connect up to 14 thermocouples. Up to 16 DBK52s may attach to a single

LogBook or Daq Device to measure a maximum of 224 (16 × 14) temperatures.

• Standard, miniature jacks can connect to all T/C types.
• Onboard cold junction sensor (CJS) circuitry is accessible via a built-in channel to the system.
• A built-in short-to-ground channel back to the system eliminates offset errors via software.
• RC low-pass input filters reject high frequency noise for thermocouple shields.
• An external analog-common access jack provides a bias-current return path for differential

signal inputs.

• An onboard programmable gain amplifier supports most thermocouple types and temperature

ranges.

• Software-based linearization (in LogView and DaqView) converts analog signals into

temperature values.

Reference Note:
In regard to Daq devices, related material for programmers is included in the separate
Programmer’s Manual (p/n 1008-0901).

The following table shows the temperature range for each thermocouple type.

T/C Type

J

K

T

E

N28

N14

S

R

B

Temperature
Range °C

-200 to
760

-200 to
1260

-200 to
400

-270 to
1000

-270 to
400

0 to
1300

0 to
1780

0 to
1780

0 to
1820

Temperature
Range °F

-328 to
1400

-328 to
2300

-328 to
752

-454 to
1832

-454 to
752

32 to
2372

32 to
3236

32 to
3236

32 to
3308

Thermocouple accuracy depends on the thermocouple itself, the cold-junction sensor, A/D conversion, and
linearization. The DBK52, LogBook, and Daq device systems use quality components and software
compensation to minimize error from these sources. Systems achieve a ±1°C CJS accuracy that contributes
to the overall accuracy. When the software is calibrated, gain and offset errors are virtually eliminated.
The DBK52’s built-in auto-zero channel dynamically eliminates any other system offset-errors.

DBK Option Cards and Modules

989594

DBK52, pg. 1

Advertising