DBK19

14-Channel Thermocouple Card

Overview …… 1
Hardware Setup …… 2

Card Connection …… 2
Card Configuration …… 3
DaqBook/100 Series & /200 Series and DaqBoard [ISA type] Configuration ……3
DaqBook/2000 Series and DaqBoard/2000 Series Configuration …… 3

Software Setup …… 3
DBK19 – Specifications …… 4


Reference Notes:

o

Chapter 2 includes pinouts for P1, P2, P3, and P4. Refer to pinouts applicable to your

system, as needed.

o

In regard to calculating system power requirements, refer to DBK Basics located near the

front of this manual.

o

Refer to the Signal Management chapter for information regarding Two-Point

Calibration of a Temperature Measurement System.

Overview

DaqView Users: When a DBK19 is used with a /2000 Series Device, the Internal Clock Speed
should be set to 100 kHz as described in chapter 3, DBK Setup in DaqView.

The DBK19 can attach to 14 thermocouples (T/C) for temperature measurements. The DBK19 features an
on-board cold junction sensor and a programmable-gain amplifier for measuring type J, K, T, E, N28, N14,
S, R, or B thermocouples. The table shows the temperature range for each thermocouple type.

T/C Type

J

K

T

E

N28

N14

S

R

B

Temperature
Range °C

-200 to
760

-200 to
1260

-200 to
400

-270 to
1000

-270 to
400

0 to
1300

0 to
1780

0 to
1780

0 to
1820

Temperature
Range °F

-328 to
1400

-328 to
2300

-328 to
752

-454 to
1832

-454 to
752

32 to
2372

32 to
3236

32 to
3236

32 to
3308

True differential measurements of thermocouples require a bias current referenced to analog common;
therefore, resistors connect each input terminal to the analog common. Filter capacitors across each input
operate with input protection resistors to form single-pole RC low-pass filters.

Each DBK19 includes a cold-junction compensation circuit wired to channel 0. Channels 2 through 15
accept external thermocouples. The DBK19s can be used with other analog expansion cards and feature
screw terminals for direct connection. Up to 16 DBK19s can be attached to a single LogBook or a Daq
Device to measure up to 224 (16 × 14) temperatures.

DBK Option Cards and Modules

879795

DBK19, pg. 1