Test all good parts …… 2-4, Test all “pass” (good) parts – Measurement Computing eZ-NDT version 6.0.9 User Manual

Page 11

Advertising
background image

eZ-NDT

938691

Investigation Mode Testing 2-5

9.

Test all

“Pass” (Good) Parts.

Investigation Mode, Pass Part Template

To test all “Pass” (Good) parts:

a) Select the Pass parts in the grid (see figure, upper right). Note that you can right-click on a part

number, e.g., “1” to access its associated template.

b) Strike the parts several times at a consistent location with a consistent amount of force.

c) If you do not see anything on your plot, click on the <Autoscale> button, or the <Log Scale Y Axis>

button, to expand the y-axis.

Autoscale Log Scale Y Axis

Use of the logarithmic scale (Log Scale Y Axis button) tends to enhance the peaks. This makes
it easier determine significant peaks.

Linear Scale of Initial Strike

d. Click the Autoscale button to have the Y axis minimum and maximum scaled to the data

Advertising