SL Power Electronics GPFM1200 User Manual

SL Power Electronics Equipment

Advertising
background image

CONDOR DC POWER SUPPLIES INC.
2311 STATHAM PKWY
OXNARD, CA 93033 + 805-486-4565

RATINGS:

Input: 100-240 V ac, 18.0 A, 50/60 Hz
Output: +48 V dc/25 A, +12 V dc/0.3 A (Aux)

Notes: 1. Maximum ambient temperature for rated output is 50°C.

2. Auxiliary +12 V output may be used to power external logic and/or fan.

3. Maximum Operating Relative Humidity 96%, no condensation.

4. Storage: -40 to +85 °C. Units should be allowed to warm-up under non-condensing

conditions before application of power.

CERTIFICATION: All models are Certified to be in compliance with the applicable requirements of
UL 2601-1, 2

nd

Ed (1997), CAN/CSA 22.2 No. 601.1-M90, EN 60601-1: A2 (1995), and IEC 60601-1:

A2 (1995).

CLASSIFICATION:

(5.1) Protection against electric shock = Class I

(In accordance with sub-

clause 5 of IEC 601-1)

(5.2) Degree of protection against electric shock = Not acceptable for applied
part without additional isolation (contact factory for details)

(5.3) Protection against harmful ingress of water = Ordinary (no protection)

(5.5) Have not been evaluated for use in the presence of a flammable
anaesthetic mixture with air, oxygen, or nitrous oxide. This evaluation is to be
made on the end equipment by the OEM.

(5.6) Mode of operation = Continuous

SAFETY DECLARATION: Condor DC Power Supplies, Inc. declares under our
sole responsibility that all models listed above are in conformity with the applicable
requirements of EN 60950 following the provisions of the Low Voltage Directive
73/23/EEC.

GROUNDING: Protection Class I requires that the Ground terminal be bonded to Protective Earth in
the end application. Using this terminal for the end-product protective earthing terminal is not
recommended. A separate dedicated protective earthing point should be used.

OUTPUTS: It is recommended that the output common or return be connected to Protective Earth in
the end application. The output is not acceptable for patient connection without additional isolation.
The DC output is SELV under normal and single fault conditions. The output presents an Energy
Hazard.

OVERVOLTAGE PROTECTION: The output is monitored for an overvoltage condition. In some
applications where an overvoltage condition could result in a hazard as defined in applicable safety
standards, redundant or additional overvoltage protection may be required. Consult factory for details.

CAUTION: When performing Dielectric Strength Tests, catastrophic failure of the unit may result if a
Dielectric Strength test voltage greater than 1800 V ac is applied between primary and secondary
circuits. The components providing isolation from primary to secondary cannot be tested while installed
in the power supply without overstressing basic (primary to ground) insulation. All isolating
components are individually 100 % tested at 4800 V ac prior to installation.

41-35221-0001 Rev. A 12/13/01

Page 1 of 2

GPFM1200-48 INSTALLATION INSTRUCTIONS

GPFM1200-48

INSTALLATION INSTRUCTIONS

Advertising