Super Systems 9120 RPSC User Manual

Page 12

Advertising
background image

Model 9120 Redundant Probe System - Color

4580 – Redundant Probe System – Color

Page 12 of 23

2400

19200

76800

4800

28800

115200

9600

38400


Slave 2 Mode
This will allow the user to set the Slave 2 mode.

Note – This is configurable, but currently, it is not

used with the RPS

. The options are: MMI, Modbus, ADAM, SSi Analog Input Board, or

Yokogawa.

IP Address

This option will allow the user to modify the IP
address of the

9120 Instrument

only. The address

must be supplied in a 999.999.999.999 format, or it
will not be accepted. Contact Super Systems at 800-
666-4330 or your IT department before changing any
of the IP addresses.

Subnet Mask
This option will allow the user to modify the IP subnet
mask of the

9120 Instrument

only. The address must

be supplied in a 999.999.999.999 format, or it will not
be accepted. Contact Super Systems at 800-666-
4330 or your IT department before changing any of
the IP addresses.

Gateway
This option will allow the user to modify the IP gateway of the

9120 Instrument

only. The address must be

supplied in a 999.999.999.999 format, or it will not be accepted. Contact Super Systems at 800-666-4330
or your IT department before changing any of the IP addresses.

Analog Input Setup

The 9120 controller has two analog inputs. Each of
the inputs comes with a factory default
configuration dependent on the application (refer to
PVT type under the

Furnace Setup

section). It can

be modified prior to shipment to your facility or in
the field by a technician or qualified/trained person
with the proper security code.
Analog Input Terminals
Analog Input 1 – terminals 31 and 32
Analog Input 2 – terminals 29 and 30

For the RPS, the default input type is 2.5
Volts, with a range of 2500

.

Input Type
The thermocouple type for most applications can be
modified depending on your specific needs. Please note that in some applications, some of the inputs DO
NOT allow the user to modify the Input type.

Note: Before changing the input type, make sure to set the

appropriate jumpers, if necessary

.

The jumper will need to be manually changed on the input board before

changing the input type to a 10:1 setting (non-thermocouple types)

. To change the Input type, first select

which input you want to change by selecting it in the pull-down menu at the top of the screen. Once
selected, click

OK

and the displayed Input type under Value will be the current type.