Logger software properly scales the raw voltage so that an accurate temperature can be

calculated. The Voltage Calibration consists of two parts: a Zero Calibration and a Span

Calibration. A Zero Calibration ensures that the raw voltage is scaled properly when the actual

voltage at a thermocouple is zero volts. A Span Calibration ensures that the raw voltage is

scaled properly when the actual voltage is 90% of the upper limit of the desired range (for

example, 72 mV when the upper limit of the range is 80 mV).

Figure 39 shows what the Voltage Calibration screen looks like.

Figure 39 - Voltage Calibration screen

The horizontal row of numbers represents each circuit board containing five inputs. The

vertical column of numbers represents each individual input. A 20-input SDS Data

Logger will feature four circuit boards depicted on this screen, while a 40-input unit will

feature eight circuit boards. Since each board has five inputs, a Voltage Calibration

usually involves five inputs on the same board at one time. A single input can be

calibrated if it is suspected that the temperature reading associated with one input is

incorrect.

Range Value: The mV value shown in the drop-down box represents the range value for

the particular kind of thermocouple wire that you are using in the calibration. Pressing

the question mark (“?”) will bring up a window that shows the range value specific to

various kinds of thermocouple wire.

Zero or Span: Check “Zero” to perform a Zero Calibration. Check “Span” to perform a

Span Calibration.

All or Use Selected: Check “All” to calibrate all inputs. Check “Use Selected” to

calibrate a group of inputs selected by pressing the Select Inputs button and choosing

the inputs that you want to calibrate.