Triplett 1201 User Manual

Odel, Compact digital multimeter instruction manual

Advertising
background image

1. Introduction

The Triplett Model 1201 multimeter is a 3 1/2 digit, handheld digital multi-

meter with automatic power off and a backlit LCD display. Using a Large

Scale Integration (LSI) integrated circuit, the multimeter is designed for

stability and accuracy. The resilient outer molded cover provides additional

protection from the occasional fall, and the threaded probe tips and screw

on insulated clips provide for secure, hands-off measurements. The meter

can measure AC and DC voltage, DC current, resistance, temperature, and

forward diode voltage drop. It can also perform an audible continuity test,

and test both 1.5V and 9V batteries.

2. Product Features

1) Test lead holder.

2) Continuity test indicator (red LED).

3) LCD display with (Low Battery) and DH (Hold) annunciators.

4)

POWER Button: Press the Power Button to turn the unit on or off.

5)

LIGHT Button: Press the “LIGHT” button to turn on the LCD backlight.

The backlight will be automatically turned off after approximately 30

seconds.

5)

DISPLAY HOLD Button: Press the “HOLD” key to lock the present value

on the display. The “

DH” sign will appear on the display while the hold

function is active. Press it again to exit and resume normal display

updating.

7) Rotary Switch: Use this switch to select test functions and ranges.

8)

COM: COM and Temperature “-” Input Jack.

9)

10A: 10A Input Jack, unfused.

10)

VΩmA: V/mA/BATT/

and Temperature “+” Input Jack, fused at

200mA, 500V

11) Meter case

12) Protective molded cover

3. Safety Information

3-1 The meter is designed according to IEC-1010 concerning electronic

measuring instruments with an over-voltage category level of 600V

(CAT III) and pollution level 2.

DO NOT USE THIS METER FOR CAT

IV APPLICATIONS.

3-2 Follow all safety and operating instructions to ensure that the meter is

used safely and is kept in good operating condition.

3-3 Safety symbols:

Important safety information, refer to the operating manual.

Dangerous voltage may be present.

Double insulation (protection Class II)

4. Cautions and Warnings

4-1 To maintain product safety, use the provided test leads. Damaged test

leads should be replaced with the same model or test leads with the

same electrical specifications.

4-2 Do not use with the battery cover removed.

4-3 Always set the rotary switch to the correct position for measurement.

4-4 Do not exceed the input limits.

4-5 Do not change the rotary switch setting when the leads are connected to

an energized circuit.

4-6 Use caution when measuring voltages higher than 60V DC or 30V AC.

4-7 Replace fuse with fuse of the same rating.

4-8 To conserve battery life, always turn the meter off when measurements

are completed.

5. General Characteristics

5-1 Max Voltage between input terminal and Earth Ground: CAT III 600V

5-2 Over-range Indication: display “1” or “-1” for the most significant digit

5-3 Low Battery Indication: is displayed

5-4 Maximum LCD Display: 1999 (3 ½ digits)

5-5 Manual range selection

5-6 Auto Power Off: The meter will switch to standby mode when it has

been on for approximately 20 minutes. Press the “

POWER” button twice

to power on the meter again

5-7 Fuse protection: Fast acting 200mA/500V

5-8 Power supply: 9V battery (6F22 or NEDA 1604)

5-9 Operating Temperature: 0°C to 40°C (relative humidity <85%)

5-10 Storage Temperature: -10°C to 50°C (relative humidity <85%)

5-11 Dimension: 150x106x36mm

5-12 Weight: approx. 250g (including battery)

6. Measurement Specifications

Accuracy is specified from 18°C to 28°C (64°F to 82°F) with relative

humidity below 70%.

DC Voltage

Range

Resolution

2V

1mV

20V

10mV

200V

100mV

600V

1V

Accuracy

±(0.5% of rdg + 2 digits)

±(0.8% of rdg + 2 digits)

Impedance: 10MΩ

Overload protection: 600V DC or AC

AC Voltage

Range

Resolution

200V

100mV

600V

1V

Accuracy

±(1.5% of rdg + 5 digits)

Impedance: 4.5MΩ

Overload protection: 600V DC or AC rms

Frequency Range: 40 to 400Hz

Response: Average, calibrated equivalent to RMS of sine wave

DC Current

Range

Resolution

200μA

0.1μA

2mA

1μA

20mA

10μA

200mA

100μA

10A

10mA

Accuracy

±(1.2% of rdg + 2 digits)

±(2.0% of rdg + 3 digits)

Overload protection: Fast Acting 200mA / 500V fuse

10A input is not fuse protected. Apply no more than 10A for less than 10 seconds

Resistance

Range

Resolution

200Ω

0.1Ω

2kΩ

20kΩ

10Ω

200kΩ

100Ω

2MΩ

1kΩ

Overload protection: 500V AC / DC

Temperature

Accuracy

Range

°F

-4~302°F

± ( 5°F+ 2 digits )

302~1832°F

± ( 3% of rdg + 3 digits )

Thermocouple: NiCr-NiSi K-type (supplied)

Overload protection: Fast Acting 200mA / 500V fuse

Accuracy

±(1.0% of rdg + 2 digits)

±(1.5% of rdg + 3 digits)

±(1.0% of rdg + 3 digits)

Resolution

1°F

M

odel

1201

Compact Digital

MultiMeter

Instruction Manual

Advertising