Triplett 1401 User Manual

Odel, Compact digital multimeter instruction manual

Advertising
background image

1. Introduction

The Triplett Model 1401 multimeter is a 6000 count, handheld digital

multimeter with auto-ranging capability, AC RMS measurement, automatic

power off and a backlit LCD display. Using a Large Scale Integration (LSI)

integrated circuit, the multimeter is designed for stability and accuracy. The

resilient outer molded cover provides additional protection from the occa-

sional fall, and the threaded probe tips and screw on insulated clips provide

for secure, hands-off measurements. The meter can measure AC and DC

voltage, AC and DC current, resistance, capacitance, frequency/duty cycle,

temperature, and forward diode voltage drop. It can also perform an audible

continuity test, non-contact AC voltage detection, and test both 1.5V and 9V

batteries.

2. Product Features

1) Test lead holder.

2) CDS sensor for automatic backlight brightness control.

3) NCV Detection area (top of DMM case).

4) NCV Detection voltage detected indicator (red LED).

5) NCV Detection mode activated indicator (green LED)

6) LCD display with (Low Battery), NCV, DH (Hold) and measurement

unit annunciators.

7) SELECT button: With the rotary switch on the “

“ range, press

this button to select resistance, diode, continuity or capacitance tests.

With the switch on either voltage or current selections, this switches

between DC and AC settings. If the temperature

°C/°F is selected, it

switches between Centigrade or Fahrenheit. If the SELECT button is

held down while the meter is turned on, the “Auto Power Off ” function will

be disabled.

8) RANGE button: Press and release the “RANGE” button to enter manual

range mode. Additional presses cause the unit to change the

measurement range of the measurement function selected. Press and

holding the button for more than 2 seconds causes the meter to toggle

back to auto-ranging mode.

9) DISPLAY HOLD Button: Press the “HOLD” key to lock the present value

on the display. The “DH” sign will appear on the display while the hold

function is active. Press it again to exit and resume normal display

updating.

10) REL button: Press this button for the meter to enter relative

measurement mode. “

REL” is displayed on the LCD and the present

reading becomes the reference value. In this mode, the displayed value

is the difference between any subsequent measurement and the

reference value: Relative measurement = measurement

value-Reference value.

11) Hz/% button: On “ACV/ACA” or “Hz” range, pressing the “Hz/%” key,

switches between

Frequency or Duty Cycle measurements.

12) Rotary Switch: Use this switch to select test functions and ranges.

13) COM: COM and Temperature “-” Input Jack.

14) 10A: 10A Input Jack, fused at 10A., 500V

15) VΩmA: V/mA/BATT/

and Temperature “+” Input Jack, fused at

800mA, 500V

16) Meter case

17) Protective molded cover

3. Safety Information

3-1 The meter is designed according to IEC-1010 concerning electronic

measuring instruments with an over-voltage category level of 600V

(CAT III) and pollution level 2.

DO NOT USE THIS METER FOR CAT

IV APPLICATIONS.

3-2 Follow all safety and operating instructions to ensure that the meter is

used safely and is kept in good operating condition.

3-3 Safety symbols:

Important safety information, refer to the operating manual.

Dangerous voltage may be present.

Double insulation (protection Class II)

4. Cautions and Warnings

4-1 To maintain product safety, use the provided test leads. Damaged test

leads should be replaced with the same model or test leads with the

same electrical specifications.

4-2 Do not use with the battery cover removed.

4-3 Always set the rotary switch to the correct position for measurement.

4-4 Do not exceed the input limits.

4-5 Do not change the rotary switch setting when the leads are connected to

an energized circuit.

4-6 Use caution when measuring voltages higher than 60V DC or 30V AC.

4-7 Replace fuse with fuse of the same rating.

4-8 To conserve battery life, always turn the meter off when measurements

are completed.

5. General Characteristics

5-1 Max Voltage between input terminal and Earth Ground: CAT III 600V

5-2 Over-range Indication: display “1” or “-1” for the most significant digit

5-3 Low Battery Indication: is displayed

5-4 Maximum LCD Display: 5999 (3 ½ digits)

5-5 Auto and manual range selection

5-6 Auto Power Off: The meter will switch to standby mode when it has

been on for approximately 20 minutes.

5-7 Fuse protection: Fast acting 800mA/500V and 10A/500V fuses.

5-8 Power supply: 2 x 1.5V AA batties (R6, or LR6)

5-9 Operating Temperature: 0°C to 40°C (relative humidity <85%)

5-10 Storage Temperature: -10°C to 50°C (relative humidity <85%)

5-11 Dimension: 150x106x36mm

5-12 Weight: approx. 250g (including battery)

6. Measurement Specifications

Accuracy is specified from 18°C to 28°C (64°F to 82°F) with relative

humidity below 70%.

DC Voltage

Range

Resolution

600.0mV

0.1mV

6.000V

1mV

60.00V

10mV

600.0V

100V

600V

1V

Accuracy

±(0.5% of rdg + 2 digits)

±(0.8% of rdg + 2 digits)

Impedance: 10MΩ, >100MΩ on 600mV range

Overload protection: 600V DC or AC

AC Voltage (True RMS)

Range

Resolution

6.000V

1mV

60.00V

10mV

600.0V

100mV

600V

1V

Accuracy

±(1.0% of rdg + 3 digits)

±(1.5% of rdg + 3 digits)

Impedance: 10MΩ - Overload protection: 600V DC or AC rms

Frequency Range: 40 to 400Hz

M

odel

1401

Compact Digital

MultiMeter

Instruction Manual

DC Current

Range

Resolution

600µA

0.1µA

6000µA

1µA

60mA

10µA

600mA

100µA

6A

1mA

10A

10mA

Accuracy

±(1.2% of rdg + 2 digits)

±(2.0% of rdg + 3 digits)

-- Overload protection: 800mA/500V Fast Fuse

10A/500V Fast Fuse, 10A up to 10 seconds