Error messages – HEIDENHAIN ND 231 B v.1 User Manual

Page 12

Advertising
background image

12

Error Messages

Display

Problem

RS232 FAST

Command for measured value
output followed too quickly by
another.

1)

SIGNAL X1

Encoder signal is too weak.
The scale may be contaminated.

1)

DSR.MISSING

The connected device has not sent
a DSR signal.

1)

REF. ERR. X1

The spacing of the reference marks
as defined in P43 is not the actual
spacing.

1)

FORMAT ERR.

Data format, baud rate, etc. do not
match.

1)

FRQ.ERR. X1

Input frequency too high for encoder
input. Traversing speed may be too
fast.

1)

MEMORY ERR.

Checksum error: Check the datum,
operating parameters and compen-
sation values for nonlinear axis error
compensation. If this error recurs,
contact your service agency!

1)

These errors are important for the attached device. The
error signal (pin 19) at D-sub connection EXT is active.

Other Error Displays
If “OVERFLOW“ appears, the measured value is too large or
too small:

➤ Set a new datum.

or

➤ Move the axis back.

If all sorting signals light up, the upper sorting limit is smaller
than the lower limit:

➤ Change operating parameters P18 and/or P19.

To clear error messages:

When you have removed the cause of the error:

➤ Clear the error message with the CL key.

Error Messages

Advertising