1 oper ation – HEIDENHAIN ND 1400 Addendum User Manual

Page 23

Advertising
background image

ND 1400 QUADRA-CHEK

23

2.1 Oper

ation

In the example above the tolerance range is divided into seven
sections:

For the True Position tolerance that has been applied, the range is
0.2000 mm and is divided into seven sections (=0.0286 mm for each
section) and the nominal is 0.0000 mm. The actual True Position value
is 0.0641 mm (which falls in the second section from the nominal)
hence is given 2 plus signs.

For the radius, the tolerance range is 0.2000 mm with nominal of
7.5000 mm. If you divide the range into seven sections as above, the
deviation of 0.0036 mm from nominal size falls in the middle section,
and therefore has no plus or minus signs.

Refer to "Chapter 7: Tolerances" in the User’s Guide for tolerancing
information.

Standard Report Example

+-+--------+-------------+----------+----------+----------+----------+---------+
| QC300 Feature Printout |
+-+--------+-------------+----------+----------+----------+----------+---------+
| Date: 11-25-2013

Time: 1:02:43 PM |

| Job: __________ Part: ____________ Operator: ______________ |
+-+--------+-------------+----------+----------+----------+----------+---------+
|T|Name |Actual/APD

|Nom/Zone | Dev | Min |Max/EPD |---/+++ |

+-+--------+-------------+----------+----------+----------+----------+---------+
|P|Cir 1 |X 68.1235| 68.1500|

-0.0265| | | |

| | |Y

-42.4320|

-42.4500|

0.0180| | | |

| | |Z 0.0000| 0.0000| 0.0000| | | |
| | |TP 0.0641| 0.1000| | | 0.1000|| .++ ||
| | |rad 7.5036| 7.5000| 0.0036|

7.4000| 7.6000|| . ||

| | |f 0.0000| | | | | |
+-+--------+-------------+----------+----------+----------+----------+---------+

Advertising