4. Specifications
The MSE 1000 is an advanced system for performing production integrated
measurements at high levels of precision and accuracy. The module described in these
instructions is designed for indoor use only. The MSE 1000 components shall only be
installed as described in these instructions. Mounting, installation and maintenance are to
be performed by qualified personnel only.
MSE 1201
ID 747501-01
MSE 1201
ID 747501-02
Power input
AC 100 V ... 240 V (±10 %)
50 Hz ... 60 Hz (±2 %)
(max. 108 W)
Rated power output
50.4 W
Replaceable fuse
T 1 A / 250 V, 5 mm x 20 mm Qty. 2
Data transfer
Standard Ethernet, IEEE 802.3
Addressing
Fixed IP address or DHCP
Operating temperature
0 °C ... 45 °C
Storage temperature
-20 °C ... 70 °C
Relative humidity
80 %
Altitude
2000 m
Degree of protection
IP40
IP65
Overvoltage category
II, intended to be supplied from the building wiring
Pollution degree
2
Mass
720 g
960 g