Technical specifications, Dvs205 multi-variable sensor – Emerson Process Management DVS205 User Manual

Page 2

Advertising
background image

DVS205

Page

2

Technical Specifications

DVS205 Multi-Variable Sensor

Differential Pressure Input

Range

1

0 to 62.2 kPa (0 to 250" H2O)
0 to 248.8 kPa (0 to 1000" H2O)

Turndowns from 1:1
to 10:1 of URL

±0.10% of span

Reference Accuracy

205E

Turndowns from 10:1
to 100:1 of URL

Not allowed

Turndowns from 1:1
to 10:1 of URL

±0.075% of span

205P

Turndowns from 10:1
to 100:1 of URL

±[0.025 + 0.005(URL/Span)]% of span

Stability

±0.125% of URL for five years, with up to ±50°F (28°C) ambient temperature changes
and up to 1000 psi (68,9 bar) line pressure

Ambient Temperature Effect
per 50°F (28°C)

Spans from 1:1 to
30:1

±(0.025% URL + 0.125% of span)

Spans from 30:1
to 100:1

±(0.035% URL + 0.175% of span)

Static Pressure Effects

Zero error

±0.05% of URL per 1000 psi (68,9 bar)

Span

error

±0.20% of DP Reading per 1000 psi (68,9 bar)

Over-Pressure Limit

3,626 psi (250 bar) applied on either or both sides without damage to the sensor

Burst Pressure Limit

10,065 psi (694 bar)

Static Pressure Input

Range

Either Absolute or Gauge:

0 to 5516 kPa (0 to 800 psia/psig)
0 to 25,000 kPa (0 to 3626 psia/psig)

Turndowns from 1:1
to 10:1 of URL

±0.10% of span

Reference Accuracy

205E

Turndowns from 10:1
to 100:1 of URL

Not allowed

Turndowns from 1:1
to 10:1 of URL

±0.075% of span

205P

Turndowns from 10:1
to 100:1 of URL

±[0.03 + 0.0075(URL/Span)]% of span

Stability

±0.125% of URL for five years, with up to ±50°F (28°C) ambient temperature changes

Ambient Temperature Effect
per 50°F (28°C)

Spans from 1:1 to
30:1

±(0.05% URL + 0.125% of span)

Spans from 30:1
to 100:1

±(0.06% URL + 0.175% of span)

Over-Pressure Limit

Same as URL

1. Consult factory for special ranges and materials which may be available. For example, 0 to 6.22 kPa (0 to 25" H2O) at

±0.10% reference accuracy.

Advertising