Specification, Operation – Sealey TA102 User Manual

Page 2

Advertising
background image

3. SPECIFICATION

Range Resolution Accuracy

200mV 100µV

± (0.5% + 5)

2V 1mV

20V

10mV

± (0.8% + 5)

200V 100mV

1000V

1V

± (1.0% + 5)

Range Resolution Accuracy

20mA 10µA

± (1.8% + 3)

200mA 100µA

10A

10mA

± (3.0% + 7)

DC voltage

Input Impedance: 1MΩ for all ranges.

Overload protection: 1000V DC/AC rms

(for 200mV range) 250V AC rms

DC Current

Overload Protection: F 250mA L 250V (range 10A unfused).

Maximum Input Current: 10A (less than 15 seconds).

Maximum Voltage Drop: 200mV.

Indication: Average (rMS of sine wave).

Range Resolution Accuracy

200Ω

0.1Ω

± (1.0% + 5)

2kΩ

1Ω

20kΩ

10Ω

± (0.8% + 5)

200kΩ

100Ω

2MΩ

1kΩ

20MΩ 10kΩ

± (2.0% + 7)

Resistance

Range

Resolution

Accuracy

20kHz

10Hz

± (1.5% + 5)

Frequency

Range Resolution Accuracy

-20°C~750°C 1°C -20~0°C(-4°F~32°F)±(6.0% +6)

0~400°C(32°F~752°F)±(1.0% +7)

-4°F~1382°F 1°F 400~750°C(752°F~1400°F)±(2.0%+7)

Temperature

Range Scope (RPm) Resolution Accuracy

rPM

180~10000 1*10rPM ± (2.5% + 5)

Tachometer

Description

test condition

Cylinder Range

Resolution Accuracy

3CYL

0~120.0°

4CYL 0~90.0°

5CYL

0~72.0° 0.1°

± (2.5% + 5)

6CYL 0~60.0°

8CYL 0~45.0°

Dwell Angle

Diode test

Range

hFE Test Current Test voltage

PNP and NPN 0~1000 Ib=10µA Vce = 3V

Transistor hFE test

Audible Continuity Test

Range Resolution Accuracy

200V

100mV

± (1.0% + 5)

750V

1V

± (1.2% + 5)

AC voltage

Input Impedance: About 450kΩ for all ranges.

Frequency range: 40Hz~400Hz

Indication: Average (rMS of sine wave).

Overload protection: 1000V DC/AC rms

If the resistance of the circuit

under test is lower than 30Ω,

the audible warning will sound.

Open circuit voltage is

approximately 3V.

Description

test condition

the approximate forward

voltage of the diode under test

will be displayed on the LCD.

the forward DC current is

approximately 1mA, the

reversed DC voltage is

approximately 3V.

4. OPERATION

WARNING! Ensure that you read, understand and apply the safety and operational instructions before connecting the analyser. Only when you are sure

that you understand the procedures is it safe to proceed with testing.

WARNING! risk of electrocution. High voltage circuits, both AC and DC are very dangerous and should be measured with great care.

Operating temperature range 0°C to 40°C.

Remember to turn off the analyser when measurement is completed.

NOTE: If '1' appears in the display during a measurement, the value exceeds the range you have selected. Select a higher range.

NOTE: On some low AC and DC ranges, with the test leads not connected to a device, the display may show a random fluctuating reading. this is

normal and is caused by the high input sensitivity. the reading will stabilise and give a correct measurement when connected to a circuit.

4.1.

Data Hold Button

4.1.1. the data hold function allows the analyser to freeze a measurement reading for later reference.

4.1.2. Press the data hold button once to freeze the reading in the display. the indicator 'H' will appear in the display.

4.1.3. Press the data hold button again to return to normal operation.

4.2.

AC voltage measurement

4.2.1.

Insert the black test lead into the negative COM socket and the red test lead into the positive

socket.

4.2.2. turn the function dial to the appropriate AC voltage setting as required and switch the analyser on (see ranges above). If the voltage range to be

measured is not known, set to the highest range setting, and then select to the correct range when first reading is taken, until a satisfactory resolution is

obtained.

4.2.3. Connect the test leads to the circuit under test and read the voltage on the display.

4.3.

DC voltage measurement

4.3.1.

Insert the black test lead into the negative COM socket and the red test lead into the positive socket.

4.3.2. turn the function dial to the appropriate DC voltage setting as required and switch the analyser on (see ranges above). If the voltage range to be

measured is not known, set to the highest range setting, and then select to the correct range when first reading is taken, until a satisfactory resolution is

obtained.

4.3.3. Connect the test leads to the circuit under test and read the voltage on the display.

4.4.

DC Current measurement

WARNING! Do not make current measurements at 10A for longer than 15 seconds in every 15 minutes. Exceeding this may cause damage to

the analyser and test leads.

4.4.1. Insert the black test lead into the negative COM socket, and the red test lead into the

positive mA socket (or the positive 10A socket for currents from

200mA

to

10A).

4.4.2. turn the function dial to the appropriate setting as required (see ranges above). If the current range to be measured is not known, set to the highest

range setting, and then select to the correct range when the first reading is taken, until satisfactory resolution is obtained.

4.4.3. Connect the test leads in series with the circuit under test and read the current on the display.

4.5.

Diode measurement

4.5.1. Insert the black test lead into the negative COM socket and the red test lead into the positive socket.

4.5.2. turn the function dial to the position and switch the analyser on.

4.5.3. Connect the red test lead to the anode of the diode, and the black test lead to the cathode of the diode.

4.5.4. the approximate forward voltage drop of the diode will be displayed. If the connection is reversed, a fig “1” will be displayed.

Open circuit voltage: less than 2.8V

Overload protection: 250V DC/AC rms

© Jack Sealey Limited

tA102 Issue No: 4(I) - 16/06/14

Original Language Version

Advertising
This manual is related to the following products: