Alarm conditions – Precision Medical PM4351 EasyFlow5 User Manual

Page 7

Advertising
background image

6

ALARM CONDITIONS

On
On
Off

Oxygen level is below 83%, continue to use but contact your

provider.

On
On
On

Oxygen level is too low; turn the concentrator “OFF”, connect to

another source of Oxygen and contact your provider.

On
Flashing
Off

Flow is lower than expected; clear blocked tubing or cannula.

On
Flashing
On

Power ON, but temperature is too high to start compressor; wait 4

minutes for fan to cool device.

On
Flashing
On

Waited 4 minutes after power ON, but temperature is still too high

to start compressor; turn off power, clear blocked vents, wait and try

again.

or

Device was running, but temperature was too high during run; turn

off power, clear blocked vents, wait and try again.

On
Off
On

Power ON, but pressure is too high to start compressor;

wait 4 minutes to bleed pressure.

On
Off
On

Waited 4 minutes after power ON, but pressure was too high to start

compressor; turn off power, wait and try again.

or

Device was running, but pressure was abnormal during run; turn off

power, wait and try again.

On
Off
Off

Power ON, but AC voltage is too low for normal operation; unit will

start when AC voltage is okay.

or

Device was running, but AC voltage dropped too low for normal

operation; unit will restart when AC voltage is okay.

Off
Off
Off

Power switch is ON but no AC power is present; check if power cord

is connected.

NOTE: If alarm conditions continue, contact your provider.