TECHNISCHE DATEN

Inlet Pressure Range:

300 - 3000 psi (2068 - 20684 kPa)

Pressure Gauge Accuracy:

3 - 2 - 3% of full scale

Dimensions: (Are approximate and may vary by model)

No Gauge Models

Gauge Models

Weight:

10 oz (283 g)

12.6 oz (357 g)

Overall Length:

2.5 in (6.4 cm)

2.5 in (6.4 cm)

Width:

1.9 in (4.9 cm)

2.8 in (7.0 cm)

Height:

2.9 in (7.5 cm)

3.2 in (8.1 cm)

Pulse Settings:

1, 2, 3, 4, & 5 l/min Equivalents

Flow Settings:

2 l/min Continuous (195405)
1 through 6 l/min Continuous (195412)

Accuracy: Continuous Settings within ± 10%

PULSE ONL

Y Accuracy:

Pulse Settings: Within +/- 15% of the nominal

bolus value (at each breath rate)

Savings Ratio:

Up to 5.7:1

Trigger Method:

Inspiratory effort (Negative pressure from

patient inhalation)

Breathing Frequency:

Up to 35 b/min

Cannula Requirement:

Maximum 7 foot (2.1 m) long standard adult

single lumen nasal cannula.

Operating Conditions:
Temperature:

35°F to 105°F (1.7°C to 40.6°C)

Altitude:

Sea level to 10,000 ft (0 to 3,048 m)

Storage Conditions:
Temperature:

-40°F to 140°F (-40°C to 60°C)

Maximum Humidity:

95% Noncondensing

Oxygen Cylinder Connection: CGA 540 Valve

Ignition and Fault Tolerance: Meets ASTM G175-03

Specifications are subject to change without prior notice. Manuals available on our Website; www.precisionmedical.com