FALL TIME (RAMP DOWN)

When the output is disabled, the time for the output to go from the programmed voltage to 0 is
referred to as the ramp down (fall) time, and is adjustable from 00.0 to 99.9 seconds. The time
setting affects the slope of the decrease in output voltage.

1. Enter the menu (see PAR. 3.2.2 and Table 3-3) and rotate the encoder until Voltage display

shows

(R.DNT) to access RAMP DOWN setting.

2. Press ENTER key to view setting.

Ramp time displayed as

(0.00) in Current display.

3. Rotate encoder until Current display shows desired setting, then press ENTER to confirm.

3.3.12

MEMORY FLAG

There are 16 memory locations that can be used to store and recall power supply settings.
MEM.F is used to limit the range of locations that can be accessed. For example, when the
MEMORY FLAG is set as 05, it means that RECALL can only be executed from memory loca-
tions 01 through 05.

1. Enter the menu (see PAR. 3.2.2 and Table 3-3) and rotate the encoder until Voltage display

shows

(MEM.F) to access MEMORY FLAG setting.

2. Press ENTER key to view setting.

Memory flag displayed as

(01) in Current display.

3. Rotate encoder until Current display shows desired setting, then press ENTER to confirm.

3.3.13

BEEP

The BEEP function enables or disables the audible beep each time a front panel key is pressed.

1. Enter the menu (see PAR. 3.2.2 and Table 3-3) and rotate the encoder until Voltage display

shows

(BEEP) to access BEEP function.

2. Press ENTER key to view setting.

Beep ON (default) displayed as

(ON) in Current display.

Beep OFF displayed as

(OFF) in Current display.

3. Rotate encoder until Current display shows desired setting, then press ENTER to confirm.