4 calibration exit (local), Calibration exit (local) -4, R. 4.3.4) – KEPCO KLP Series (older -1200 models) Operator Manual User Manual

Page 76: R. 4.3.4

Advertising
background image

4-4

KLP 073008

1. The status display reads INPV. Refer to PAR 3.7 to connect the reference voltage used to

define full scale output voltage. as the source as described in PAR 3.7.

2. When the reference voltage is at the desired value, tap the DC OUTPUT switch to accept the

value. When the calibration is saved, this analog voltage will be equivalent to full scale out-
put voltage. The full scale output voltage is determined by the virtual settings in effect.

3. The status display reads INPC. Connect the reference voltage used to define full scale out-

put current as described in PAR 3.7.

4. When the reference voltage is at the desired value, tap the DC OUTPUT switch to accept the

value. When the calibration is saved, this voltage value will be equivalent to full scale output
current. The full scale output current is determined by the virtual settings in effect.

5. The status display reads EXIT. Proceed to PAR. 4.3.4 to exit calibration.

4.3.4

CALIBRATION EXIT (LOCAL)

NOTE: using a thin tool (e.g., a paper clip), tap the FUNCTION switch to exit calibration WITH-

OUT saving any calibration data. If no calibrations were completed, then the unit
advances to the next function as if the FUNCTION switch was pressed.

1. When the status display reads EXIT, tap the DC OUTPUT switch to save the calibration.

2. The status display reads DATE. The DC VOLTS display represents MMDD (month, day) and

the DC AMPS display represents YYYY (year). The factory calibration date is displayed with
the first digit blinking as a starting point to enter the current date. Rotate either the VOLT-
AGE
or CURRENT controls to change the digit, then tap either control to move to the next
digit.

NOTE: For purposes of traceability the user is strongly urged to enter the calibration date upon

completion of calibration. If a date is not entered, no date is saved.

3. Tap DC OUTPUT switch to accept date. If any calibration (voltage, current or external) was

completed, the unit will save the new calibration. The status display reads "DONE" while cal-
ibration is being saved.

4. Once saved the unit enters Virtual Model mode. Follow the procedure in PAR 3.2.5 to set the

virtual model.

NOTE: The virtual model is always reset following calibration. If voltage calibration is com-

pleted (regardless of whether or not current calibration is completed), the virtual model
voltage is set for maximum rated voltage, and current is set to the value corresponding
to 1200 Watts. If ONLY current calibration is completed, the virtual model current will
be set for maximum rated current and voltage is set to the value corresponding to 1200
Watts.

4.4

CALIBRATION USING VISA DRIVER DEMO PROGRAM SOFT PANEL

For voltage calibration all loads must be disconnected and the sense terminals connected to the
corresponding output terminals. The digital voltmeter will be connected to the output of the
power supply. For current calibration after disconnecting all loads an appropriate shunt resistor
will be connected across the output terminals and the digital voltmeter will be connected across
the sense terminals of the shunt resistor.