Amprobe DM-CT-100 Current-Transducer User Manual

Dm-ct-100, Operating instructions, S1 (+)

Advertising
background image

WARNING:

These instructions must be read in full.

All interventions on LIVE conductors are DANGEROUS.
The OPERATOR must know all the rules relating to ELECTROTECHNICAL SAFETY.
He must be trained and authorised. He is responsible for the safe use of equipment.
He must be aware:
-

Whether this equipment is appropriate for the use he wishes to make of it and whether it is in a safe condition

-

The conditions under which actions are taken.


PROBE conforms to safety STANDARD: CEI 1010-2-032.

NORMAL CONDITIONS:

inside a building, maximum altitude 2000 m, temperature of between 5ºC and 40ºC (41°F and 104°F), max. relative humidity of 80% up to 31ºC

(89°F), decreasing to 50% at 40ºC (104°F) on bare conductors with a voltage of up to 600 V CAT III. Pollution Degree 2. Insulated conductor: insulation limit of the cable.

PROBE conforms to European EMC STANDARDS -

immunity: EN 50082-2:1992 10V/m industrial environment. - Emission: EN 50081-2: 1993 Industrial environment.

USAGE:

Amprobe recorders should be set to 100A current range. Refer to individual recorder manual on how to configure current input. Verify recorder compatibility with

Amprobe catalog or a website: www.Amprobe.com

SAFETY DEVICES:

The PROBE must only be handled by part [A] (see drawing). No part of the hand must exceed the TACTILE SAFETY BARRIER (G).

This probe is fitted with a SIAC Safety Internal Anti Short Circuit (patented). To open the jaws, push the slider (V) and depress, on closure the slider will retract automatically
(make sure of it)

if no object (3 mm) is between the open jaws.

Stop using if any part of the PROBE appears to be damaged or in poor condition. Inspect, or have inspected, the condition of this equipment at regular intervals.

MAINTENANCE:

Clean the probe from time to time by wiping with a damp cloth and a detergent. Do not use abrasive products or solvents. Do not immerse the probe in

GENERAL DATA
Maximum conductor size …………..………………: 15mm diameter, bus bar 15x17mm
Operating/ Storage Temperature ........................ : -10ºC to +55ºC / -20ºC to +70ºC (14°F to 131°F/-4°F to 158°F).
Enclosure Protection .......................................... : IP20 (EN 60529)
Weight ................................................................ : 175 g
Output Connection ............................................. : 2m lead with D01 connector
Overheating will occur if the product is subjected to : (Ip rms x frequency) > 1,000,000. Limit measurement time.
Power measurement: Respect the conventional P1=S1 direction of the arrow on the probe. (see drawing)

Primary Current

0.5 to 10 A

10 to 50 A

50 to 100 A

100 to 150A

Accuracy (% of reading) ± 3% +5mV

± 2% +5mV

± 1% +5mV

± 1% +5mV

Phase Shift

Not Specified

2.6°

International Electrical Symbols

Caution! Refer to this manual before using the probe

Probe is protected by reinforced or Double Insulation

bla DM-CT-100_En_Sp_060405.pub

Operating Instructions

DM-CT-100

LIMITED WARRANTY
Congratulations! Your new instrument has been quality crafted according to quality standards and contains quality components and workmanship. It has been inspected for
proper operation of all of its functions and tested by qualified factory technicians according to the long-established standards of our company. Your instrument has a limited
warranty against defective materials and/or workmanship for one year from the date of purchase provided that, in the opinion of the factory, the instrument has not been
tampered with or taken apart. Should your instrument fail due to defective materials, and/or workmanship during this one year period, a no charge repair or replacement will be
made to the original purchaser. Please have your dated bill of sale, which must identify the instrument model number and serial number and call the number listed below:

Repair Department
ATP – Amprobe
Phone: 954-499-5400
Toll Free: 800-327-5060
Fax: 866-287-7222
Website: www.Amprobe.com


Please obtain an RMA number before returning product for repair. Outside the U.S.A. the local representative will assist you. Above limited warranty covers repair and
replacement of instrument only and no other obligation is stated or implied.

REFERENCE CONDITIONS:

Temperature: +19°C to +25°C (66°F to 77°F) ;Humidity: 20-75% RH ; Sinusoidal current: 48-65 Hz ; Distortion factor: < 1% ;DC Current: None ;

DC magnetic field: Earth’s Magnetic Field (< 40A/m) ; Alternating magnetic field: None ; Proximity of external conductor: No current ; Position of conductor: Centered between
jaws.

Primary nominal current IN

100

ARMS

Conductor position sensitivity

< 0.5% at 50/60 Hz

Measuring range

0.5 A to 150

ARMS

Error due to adjacent conductor

<15 mA/A at 50 Hz

Output sensitivity

10 mV/A

Phase shift at 2 kHz & 10A: ± 1º

± 1º

Load impedance

≥ 100 kΩ

Temperature coefficient

0.015% / ºC

Frequency range

40 Hz to 10 kHz (-3 dB)

ELECTRICAL CHARACTERISTICS

G

A

NM C

3

V

"SIAC"

CLP

-

1

NC

COM

S1 (+)

D01