• This crowbar is activated by voltage. If the voltage is too high, that

crowbar will blow the fuse automatically to interrupt the circuit.

• The voltage between P1 and P2 is also protected for over voltage but

does not have a fuse.

• The tester has a temperature shut down. The temperature sensing is

done on the current regulation transistor.

• Should this over-temperature LED lit, allow the instrument to cool down

for a while before proceeding further.

• The milliohm meter, with its measuring range of 100u ohms to 2000

ohms, is suitable for a wide range of applications such as:

1. Measuring the winding resistance of electric motors, generators and

transformers.

2. Bond testing in mines, aircraft, railways, ships, domestic andindustrial

wiring installations.

3. Measuring the ring main continuity testing in industrial and domestic

wiring installations

4. Measuring resistance in electronic equipment such as shunts, pcb

tracks, switch and relay resistance.

5. Checking compression joints on overheads lines.

6. Testing and maintenance of switchboard /sub-station equipment on

such items as fuses, joints, contacts and bonds.