Data Memory – Saving individual measurements

This functions allows saving maximum 99 individual measurements.

A) Saving measurements

1) Switch on the instrument
2) Perform the measurement
3)

Save the desired measurement pressing the “M/MEMORY” key (19).

When saving a measurement, the symbol [M] is briefly displayed, the storage

place is indicated below.

B) Read out the measurement

1)

Press the “R/READ” key (20). The symbol [R] (5.43) is displayed.

2) The navigation keys “5“ und “6” (16) allow to select and display the
measurements, starting with the first storage place.
3)

Press the key

(16a) to leave the memory mode again.

C) Delete measurements

1) Switch off the instrument
2)

Press the “M/MEMORY” key (19) and keep it pressed.

3) Switch on the instrument.
4)

Keep the “M/MEMORY” key pressed until the symbol “CLr“ is briefly
indicated in the lower right corner of the display.

Recording with the Data Logger

This functions allows to record up to 21,000 measurement values by means of the
internal data logger. For evaluation purposes, the internally recorded data must
be transferred to a PC (please refer to page 32).

To use the total data memory, delete the data memory for the data logger

before starting the recording.

Delete Measurement Data within the Data Logger

The values within the data logger may only be deleted using the software

supplied (please refer to page 32).

22