Emerson Process Management MICRO MOTION 1500 User Manual

Page 7

Advertising
background image

Environmental limits for R-Series sensors

Figure 1-1:

Maximum process temperature in °F (°C)

176 (80)

140 (60)

104 (40)

68 (20)

32 (0)

–4 (–20)

–40 (–40)

–148 (–100)

158

(70)

Ambient temperature of core processor or

transmitter in °F (°C)

–112 (–80)

–76 (–60)

–40 (–40)

–4 (–20)

32 (0)

68 (20)

104 (40)

140 (60)

176 (80)

212 (100)

248 (120)

302 (150)

104
(40)

Notes

When ambient temperature is below –40 °F (–40 °C), a core processor or transmitter must be
heated to bring its local ambient temperature to between –40 °F (–40 °C) and +140 °F (+60 °C).
Long-term storage of electronics at ambient temperatures below –40 °F (–40 °C) is not
recommended.

Temperature limits may be further restricted by hazardous area approvals. Refer to the hazardous
area approvals documentation shipped with the sensor or available from the Micro Motion web site
(

www.micromotion.com

).

The extended-electronics option allows the sensor case to be insulated without covering the
transmitter or core processor, but does not affect temperature ratings.

Planning

Installation Manual

3

Advertising