Calibration mode, Iqs-1500, The iqs-1500’s summary window – EXFO Photonic Solutions Div. IQS-1500 User Manual

Page 3

Advertising
background image

The IQS-1500’s Windows-compatible, highly intuitive GUI allows easy control via mouse, touchscreen or keyboard. Also, it lets you save multiple
user configurations and access the instrument’s online user guide.

Calibration Mode

IQS-1500

A

Start

Allows the user to select the calibration mode, start a new calibration
sequence with template file or view a report file.

B

Information

Allows the user to enter general and specific information about current
measurement conditions. This information will be included in the
calibration summary.

C

General

Contains user and module information currently used to perform
calibration.

D

Conditions

Contains all the information pertaining to the conditions at which the
calibration is performed: source wavelength, power and bandwidth,
temperature and humidity.

E

Device under Test (DUT)

Contains all the information pertaining to the DUT under calibration.

F

Uncertainties

Allows the user to enter specific information pertaining to the
calculation of the calibration uncertainty.

G

Null

Allows the user to eliminate electronic offsets and dark currents to
ensure accuracy.

H

Calibration

Allows the user to perform calibration measurements and calculation.

I

Ref. Meter Measurement

This auto-save function allows the user to select the number of
measurements and the intervals between readings for automated
measurements on the reference power meter (IQS-1500).

J

DUT Measurements

Allows the user to enter DUT power measurements.

K

Summary

Displays complete calibration results, including mean power, calibration
factor and total uncertainty.

L

Detailed calibration reports

Generates detailed calibration reports that include instrumentation
identification, environmental information, reference conditions and
a calibration summary for each wavelength.

Source wavelength

Wavelength setting of
power meter under test

Mean power measured
by the IQS-1500

Number of calibration
measurements

Calibration factor for DUT
readings (in W)

Standard deviation of the
calibration factor

Expanded uncertainty of
corrected DUT readings

1

1

5

6

7

2

3

4

2

3

4

5

6

7

The IQS-1500’s Summary Window

A
B

G
H

I

J

K
L

C
D
E
F