Troubleshooting 131

or after you rewind when practicing without a reservoir when your first get your pump. In this case,

it is just a reminder to make sure all your settings are programmed before you begin using the pump
with insulin.

My screen appears distorted

The screen may appear distorted or have a “rainbow” appearance if you are wearing polarized

sunglasses, are in bright sunlight, or in extreme high or low temperatures. If your screen appears
distorted:

Take off your sunglasses.

Move into the shade.

Make sure your pump is not in direct heat (i.e., next to a heater) or cold (worn on the outside of

your clothing on a very cold day).

Do not return the pump: this is a normal property of this type of screen on any device.

I cannot get out of the priming loop

1

Is the reservoir in the pump full?

If no, place a full reservoir or shipping cap in the pump.

If yes, make sure you are disconnected from the pump.

2

Hold the

ACT

button until the second set of beeps and the numbers appear on the screen.

If yes, your pump is okay, go to step 4 in the “Manual prime” instructions on page 49 to finish the
manual prime.

If you did not hear a second set of beeps or numbers did not appear on the screen, change your
infusion set and repeat this step.

3

If you still do not hear the beeps and see the numbers count up on the screen, call the 24-hour
Product Help Line.

The pump is asking me to rewind

This is normal after any of the following:

1

any E-alarms,

2

the Clear Settings function,

3

or, a “No Delivery” Alarm (during the Prime sequence).