Specifications and utility requirements, Erformance, Ystem – HP 11744195 User Manual

Page 9: Etails, Rogrammable, Ontrol and, Ogging, Imensions and, Tilities, Able

Advertising
background image

Chart Environmental Chambers

Specifications and Utility Requirements

Specifications and Utility Requirements

Table 1 - Chamber Performance Specifications

P e r f o r m a n c e

T e m p e r a t u r e

Allowable Range

-80 to +200

°

C

Temperature change rate

Up to 20

°

C/min

Temperature Control

+/- 2

°

C after stabilization

Table 2 - System Details

S y s t e m D e t a i l s

Cooling LN

2

direct injection, time proportional control.

Heating

2 kW (4 kW in HP model) of heat provided by open nichrome
wire heating elements. Time proportional solid state control of
resistive heating elements in air conditioning plenum.

Programmable Control and Data Logging

Controller

Watlow Series 96 temperature controller.

Communications

A serial communications connector is installed in the wall of
the control panel, wired to the Watlow controller to facilitate
communication with a PC. This feature is not standard, but
can be ordered as an option.

Table 3 - Dimensions and Utilities

D i m e n s i o n s a n d U t i l i t i e s

Interior Workspace

12” W x 14” D x 12” H

Overall Exterior Dimensions

116” W x 146” D x 124” H

Weight of System

220 pounds (Approximate)

Electrical

230 VAC 1PH; 20 FLA (HP model - 30 FLA)

Nitrogen

1/8” fpt; 40-50 psig, 175 psig MAX

Exhaust 1”

OD


4