  White calibration for reflectives 

The Flextight X5 scanner uses a white reference to make sure that all elements in the

scanner’s CCD react consistently to light brightnesses. The scanner has been calibrat-

ed at the factory, and the calibration data is stored in the scanner’s flash PROM.
As the scanner ages, the properties of the light tubes may change slightly. Therefore,

you must redo the white calibration every six months or so and immediately after

changing the light tubes.

Use the procedure below to make a White Calibration:

1. Find the white calibration sheet that came with the scanner. It is a square white

piece of paper (219 x 219 mm (8.6” x 8.6”)). If you have lost the white calibration

sheet, then you can use a flat (non-textured), perfectly white piece of paper of the

same dimensions.

2. Your scanner came with several clear plastic layers for the reflective original holder.

If you have used the reflective holder before, make sure that there are no finger

prints or other marks on the plastic layer. Replace the layer with a new one if

necessary.

3. Align the top edge of the sheet with the dotted line near the top of the original

holder (see the figure below). Align each side of the target with the edges of the

original holder.

4. Select

White Calibration from the Maintenance window. The External White

Calibration window appears.

5. Click on

Calibrate. The scanner will take several scans of the white target and the

whole process will take about eight minutes.

13