2. Specifications

2.1 Specifications for 04905 Series

Parameter Description

Cell Constant

04905 Series: 0.01, 0.1, 1.0, 10 and 50 as
specified

Electrode Material

Nickel, Platinum or Monel as specified

Maximum Pressure Limit

1724 kPa @ 140°C (250 psig @ 284°F)

Maximum Continuous Temperature Limit

140°C (284°F)

For immersion applications:
80°C (176°F)

Materials of Construction

Cell Body: PES (polyethersulfone)

Support Fittings: Ryton

Electrodes: Nickel, Platinum or Monel as specified

Quick Disconnect Receptacle: Stainless Steel

Cable Options

Leadwire: PVC insulated 22 gage cable, 0.245”
OD, 20 and 50 feet lengths available.

Quick Disconnect Option
Mating cables must be purchased from Honeywell

Universal Head (Aluminum)

Weight

Approximately 1 lb (0.45 kg)

If using universal head:
3 lb (1.35 kg)

Approvals

Manufactured to comply with ASME boiler and
pressure vessel code Section III, Div.1, UG-101

CRN #0F11607.5C

Insertion

1” NPT male, Schedule 40

Flow Chamber

Inlet: 3/4" MNPT

Outlet: 3/4" FNPT

Insertion Depth

5” to 7” (127 to 178 mm) depending on cell
constant

Overall Length

Approximately 6 to 8” (152 to 203 mm)

If using universal head:
10 to 12-1/4” (254 to 311 mm)