TROUBLESHOOTING

If any abnormality arises during use, please check the following points

Note: If the unit still does not work, contact HoMedics Consumer Relations.

Under no circumstance should you disassemble or attempt to repair
the unit by yourself. Contact information for HoMedics Consumer
Relations Department can be found on the warranty page.

SYMPTOMS

POSSIBLE

CAUSES

CORRECTION

Unit does not turn on
when the
START/STOP
button
is pushed.

Batteries have
run down.

Replace them with
four new AA
alkaline batteries.

Battery polarities
have been positioned
incorrectly.

Re-insert the batteries
in the correct positions.

EE measurement error
symbol shown on dis-
play or the blood pres-
sure value is displayed
excessively low
(or high).

Cuff has been placed
incorrectly.

Wrap the cuff properly
so that it is positioned
correctly.

Did you talk or move
during measurement?

Measure again. Keep
arm steady during
measurement.

Shaking of the arm with
the cuff on.

E1 error symbol shown
on display

Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.

Check cuff connection.
Measure again.

E2 error symbol shown
on display

Inflation pressure
exceeding 300 mmHg.

Switch the unit off, then
measure again.

E3 error symbol shown
on display

Error determining
measurement data.

Measure again.

Power Source:

Four 1.5V DC (AA) batteries or 6V DC AC adapter

Measurement Method: Oscillometric

Measurement Range:

Pressure: 40~280 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute

Accuracy:

Pressure: ±3 mmHg
Pulse: ±5% of reading

Pressure Sensor:

Semi-conductor

Inflation:

Pump driven

Deflation:

Automatic Pressure release valve

Memory Capacity:

60 memories for each User (120 total)

Auto-shut-off:

1 minute after last button operation

Operation
Environment:

Temperature: 50°F~104°F (10°C ~40°C)
Humidity: Less Than 85% RH

Storage Environment:

Temperature: -4°F~140°F (-20°C~60°C)
Humidity: Less Than 85% RH

Weight:

0.72 lb (325 g) (without batteries)

Arm Circumference:

Standard size cuff: 9"–13" (23-33 cm)
Large size cuff:13"–17" (33-43 cm)

Dimensions:

5.9"(L) x 4.4"(W) x 2.2"(H)
149mm(L) x 111.5mm(W) x 57mm(H)

Internally Powered equipment
Type BF Equipment
IPx0-No special protection against the ingress of
water
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or nitrous
oxide.
Continuous operation with short-time loading.

SPECIFICATIONS