Voltage Test with RCD Trip Test

(only T140 VDE)

1)

Press Button on rear side -

measurement point light (10).

2)

Measure between L and PE.

3)

The RCD is tripping

The RCD could be so long tripped like the

measuring point lighting is active and L is

measured against PE. After a voltage test

wait approx.. 50 seconds to trip the RCD

again.

Restistance Test (only T140 /
T140 VDE)

Make sure that UUT is not live.

• Check that UUT is not live by carrying out a

double-pole voltage test.

• Connect both test probes with UUT. Press

Button on rear side - for measurement point

light (10) and read value on the display.

The resistance range is 1…1999

1 at a

resolution of 1

1.

The resistance test is active for 20 se-

conds after having pressed the button on

the rear.

If during the resistance measurement a

voltage is present the instrument switch

automatically to voltage measurement.

Continuity Test / Diode Test

Make sure that UUT is not live. Test volta-

ge polarity at handle test probe is positive

(+).

• Check that UUT is not live by carrying out a

double-pole voltage test.

• Connect both test probes with UUT. A signal

sound is audible for continuity and the LED

for continuity Rx/

1 is illuminated.

Fluke T100/120/140

Users Manual