Symbols, Using the probe, Specifications – Fluke 80K-15 User Manual

Page 2

Advertising
background image

80K-15

Electronic Air Cleaner Probe

Symbols

T

Double insulated.

W

Refer to explanation in this instruction sheet.

X

Lethal voltages may be present.

Complies with relevant Canadian Standards
Association directives.

Earth ground

Using the Probe

87

TRUE RMS MULTIMETER

MIN MAX

RANGE

HOLD

H

Hz

REL

mA

A

V

V

OFF

!

!

A

COM V

mA

µ

A

1000V MAX

400mA MAX

FUSED

10A MAX

FUSED

PEAK MIN MAX

41/2 DIGITS

1 Second

µ

A

CAT II

III

mV

DC

5

3

4

1

2

EAOO1F.EPD

A

Probe tip

B

Probe body

C

Plug

D

GND tab

E

Grounding lead clip

F

Multimeter or voltmeter

Figure 1. 80K-15 Electronic Air Cleaner Probe

The 80K-15 (Figure 1) is designed to work with a meter that
has an input impedance of 10 M

(

±

1% ). A meter with an

input impedance less than 10 M

can produce readings

lower than the voltage present. This can pose a serious
hazard when a dangerous voltage is present.

The 80K-15 represents a 1000 M

load to the circuit being

measured, or 1

µ

A per 1 kV. Table 1 shows the circuit

loading and input/output characteristics of the probe over its
measurement range.

Taking care to follow the safety practices under “

W

Read

First: Safety Information,” use the probe as follows:

1. Turn on a compatible voltmeter or multimeter and

select the voltage measurement function.

2. Select an appropriate voltage range (i.e., 1 volt reading

per 1000 volt input. See Table 1).

3. Connect the probe’s output leads to the voltmeter or

multimeter input terminals.

4. Connect the alligator clip of the probe’s grounding lead

to earth ground.

5. Touch the probe tip to the circuit being measured and

read the measurement on the meter display. If
necessary, apply a correction factor to the reading.
(See “Note” in the “Specifications,” below.

Table 1. 80K-15 Circuit Loading and Input/Output

Characteristics

Input Voltage

Loading Current

Output Voltage

10 V

10 nA

10 mV

100 V

100 nA

100 mV

1 kV

1

µ

A

1 V

10 kV

10

µ

A

10 V

15 kV

15

µ

A

15 V

Specifications

The 80K-15 Electronic Air Cleaner Probe achieves its rated
accuracy when it is used with a voltmeter or multimeter (ac
or dc) with an input impedance of 10 M

±

1%.

The accuracy of the probe does not include the accuracy of
the meter. To get the accuracy of the system, the accuracy
of the meter must be added to the accuracy of the probe.

Note

The input impedance of the Autoranging mode on
Fluke handheld digital multimeters varies by range.

The input impedance in almost all ranges of Fluke
multimeters is ~10 M

. The exceptions are: the 3 V

range on the Fluke Models 21, 23, 25, 27, 70, 73,
75, 7; and the 4 V range on the Models 10, 11, 12,
16, 79, 83, 85, 86, 87, 88. In the ranges on these
multimeters, the impedance is 11.11 M

.

To improve the measurement accuracy of the probe
when using these ranges on the models indicated,
apply a correction factor of 0.99 i.e., multiply the
display reading by 0.99.

Voltage Range:

1 kV to 15 kV dc or peak ac,
10 kV rms ac

Maximum Current:

20

µ

A

Input Resistance:

1000 M

Division Ratio:

1000 : 1 (1000 x attenuation)

Accuracy DC:

±

2 % in 10

o

C to 45

o

C

Accuracy AC:

±

5 % @ 60 Hz, in 20

o

C to 30

o

C

Safety:

Complies with IEC 1010-2-31:1993,
Type B, 15 kV dc or peak ac, 10 kV
rms ac, Overvoltage Category I
(voltages derived from limited energy
transformer).

Altitude:

2000 meters