Specifications, Specification for clamp meter/coil interaction, Calculating total specification – Fluke 5500A User Manual

Page 2

Advertising
background image

Specifications

Number of Turns

50

Maximum Current

11A rms, continuous

20.5A rms, 2 minutes

Maximum Duty Cycle Derating

< 11A, continuous

> 11A, 2 minutes ON, 8 minutes OFF

Maximum Voltage

7 V dc or 7 V rms ac

Minimum Inner Diameter of Clamp Jaws

2.54 cm [1”]

Specification for Clamp Meter/Coil interaction

Specification for 5500A/COIL

Coil Specifications [clamp/coil interaction]

Calibrator Output

Effective

Current Output

Toroidal-type Clamps

[such as the Fluke 80i-600 and 80I-1000]

'other-type' of Clamps

[such as the Fluke 80i-KW, -400,
80i-410, -500, -1010 and 30-series]

Magnitude

Frequency

Amp-turns

±(% of output + Amps)

±(% of output + Amps)

0.2 - 0.33 A

DC

10 - 16.4999

0.50

0.02

0.33 - 2.9999 A

DC

16.5 - 149.999

0.5

0.14

3.0 - 20.5 A

DC

150 - 1025

Not applicable

0.5 0.5

0.2 - 0.33 A

45 Hz to 65 Hz

10 - 16.4999

0.28

0.003

0.56

0.03

0.33 - 2.9999 A

45 Hz to 65 Hz

16.5 - 149.999

0.28

0.025 0.56 0.25

3.0 - 20.5 A

45 Hz to 65 Hz

150 - 1025

0.28

0.09

0.56

0.9

0.2 - 0.33 A

65 Hz to 440 Hz

10 - 16.4999

0.79

0.003

1.00

0.03

0.33 - 2.9999 A

65 Hz to 440 Hz

16.5 - 149.999

0.79

0.027 1.00 0.25

3.0 - 20.5

**

A

65 Hz to 440 Hz

150 - 1025

0.79 0.1

1.00

0.9

0.2 - 20.5 A

Above 440 Hz

10 - 1025

Usable but unspecified

**Note: Some calibrators may not have adequate compliance voltage to drive coil in this range.

Calculating Total Specification

The total specification of the effective current that the clamp-type meter measures is a function of the
clamp/coil interaction and of the current calibrator. To determine the total specification use the following
formula:

source

2

coil

2

total

U

S

S

+

=

Example:

Assume we are driving the coil with the Fluke 5500A at 4A, 60 Hz (the clamp meter will see an effective
current of 200A, 60 Hz) and we are calibrating a toroidal-type clamp meter. The Calibrator’s 1 year
specification at 4 Amps is

± (0.06% + 2 mA), so the effective current in the coil bundle will have a speci-

fication of

± (0.06% + 0.1A). Next, we find the total specification of the calibrator and the coil as a per-

centage of the output:

Specification of effective calibrator current in coil bundle =

± (0.06% + 0.1A) = 0.11%

Specification due to clamp Meter/Coil Interaction =

± (0.28% + 0.09A) = 0.325%

The RSS of these two specifications determines the total specification of the clamp/source combination:

%

343

.

0

%

11

.

0

%

325

.

0

2

2

total

=

+

=

S