Overload protection for all functions and ranges: Rated voltage.

Function

Input Impedance (Nominal)

L

, m

L

,

K

>10 M

, <50 pF

Common Mode
Rejection Ratio
(1 k

Unbalanced)

Normal Mode Rejection

L

, m

L

>120 dB at dc,
50 Hz, or 60 Hz

>60 dB at 50 Hz or 60 Hz

K

>60 dB dc to 60 Hz

Open Circuit Test
Voltage

Full Scale Voltage

To 3.2 M

32 M

e

<3.1 V dc
<2.8 V dc (typical)

<440 mV dc
<420 mV dc
(typical)

<1.4 V dc
<1.3 V dc
(typical)

Open Circuit Test
Voltage

Full Scale Voltage

G

R

<3.1 V dc

2.0 V dc

Short Circuit Current (typical)

e

21/75:

400

µ

A

23/77:

300

µ

A

G

R

500

µ

A

400

µ

A

G

V

F

Current (typical)

0.0 V
0.6 V
1.2 V
2.0 V

21/75:

0.5 mA
0.4 mA
0.3 mA
0.1 mA

23/77:

0.4 mA
0.3 mA
0.2 mA
0.1 mA