Cloning a Transmitter

19

Cloning a Transmitter

Cloning copies the Basic Setup information from one
transmitter to another. You can clone generic and
supported transmitters.

To use the calibrator to clone a transmitter:

1. Connect the calibrator to the transmitter you want to

clone. You only need a communication connection.

2. From the Active Device screen, press the

Setup

softkey.

3. Use

the

u

or

d

keys to select

Basic

from the list on

the screen and press

e

.

pe09s.bmp

4. Press

the

Store Page

softkey to load the settings into

the calibrator’s memory (not the transmitter).

5. Disconnect the calibrator from the transmitter and

connect it to the transmitter you want to configure.
Press the

Abort

softkey to establish communication.

6. Press the

Setup

softkey, then select

Basic

to return to

the Basic Setup screen.

7. Press the

Recall Page

softkey to refresh the calibrator

display with the cloned parameter settings. The
settings are not transmitted to the transmitter at this
time.

8. Use the

u

or

d

and

e

keys to select

parameters one at a time, change them, or leave them
as cloned, and press the

Send

softkey to transmit

them to the transmitter.