Replacement parts and accessories – Fluke 52 Series User Manual

Page 19

Advertising
background image

51 & 52 Series II

Replacement Parts and Accessories

15

Electrical (cont.)

Measurement

Accuracy, T1,

T2, or T1-T2

(Model 52)

J-, K-, T-, and E-type:

±

[0.05

%

of

reading

+

0.3

o

C (0.5

o

F)]

[below

100

o

C (

148

o

F): add 0.15 %

of reading for J-, K-, E-, and N-type;
and 0.45 % of reading for T-type]

Temperature

Coefficient

0.01

%

of reading

+

0.03

o

C per

o

C

(0.05

o

F per

o

F) outside the

specified

+

18

o

C to 28

o

C (

+

64

o

F

to

+

82

o

F) range

[below

100

o

C (

148

o

F): add

0.04 % of reading for J-, K-, E-, and

N-type; and 0.08 % of reading for

T-type]

Electromagnetic
Compatibility

Susceptibility:

±

2

o

C (

±

3.6

o

F) for 80

MHz to 200 MHz in 1.5 V/m field, for

200 MHz to 1000 MHz in 3 V/m field.
Emmisions: Commercial Limits per
EN50081-1

Maximum

Differential

Common Mode

Voltage

1 V (Maximum voltage difference

between T1 and T2)

Temperature

Scale

ITS-90

Applicable

Standards

NIST-175

Accuracy is specified for ambient temperatures

between 18

o

C (64

o

F) and 28

o

C (82

o

F) for a period of

1 year. The above specifications do not include

thermocouple error.

Replacement Parts and Accessories

Accessory

Part Number

Holster and Flex Stand

Assembly

1272438

AA NEDA 15A IEC LR6 batteries

376756

80PK-1 K-Type Bead Thermocouple

773135

CD-ROM

1276106

Service Manual

1276123

Advertising