Disinfecting materials, Disinfection/cleaning kit 12214 – GE Z-14400 User Manual

Page 60

Advertising
background image

Z Series RO System by GE Osmonics

4.4.1

Formaldehyde Disinfection Preparation

WARNING:
Always wear rubber gloves, apron and facemask. Use proper ventilation during
disinfection and rinse-out. A face respirator with an organic cartridge to remove
formaldehyde fumes may be worn. DO NOT BREATHE FUMES OR ALLOW
FORMALDEHYDE TO COME IN CONTACT WITH SKIN OR EYES. Follow all
Material Safety Data Sheet (MSDS) precautions and OSHA Standards.

WARNING:
Label machine with appropriate warning signs such as "Do Not Use / Contains
Disinfectant" to prevent use of the RO until it is properly rinsed.

CAUTION:
New or replacement thin film membranes must be rinsed to drain with normal pressures
for a minimum of six (6) hours before first exposure to formaldehyde, or irreversible
damage may occur.

Disinfecting Materials
Description

GE Osmonics Part Number

Disinfection/Cleaning Kit

12214

Formaldehyde, 37% Solution

Not Provided by GE Osmonics

Formaldehyde Test Kit

Not Provided by GE Osmonics

“DO NOT USE” Warning Labels Not Provided by GE Osmonics
Cleaning/Disinfect Log Sheets

Reference Tech Notes 143, 144, 145

Operation Log Sheet

Reference Tech Note 142

Body Protective Wear

Not Provided by GE Osmonics

1238339b – 16Oct03

4 - 16

Routine Maintenance

Advertising
This manual is related to the following products: