Graham Field LABTRON 707AC User Manual

Page 7

Advertising
background image

Labtron Digital Blood Pressure Monitor

14

Labtron Digital Blood Pressure Monitor

7

tRoUBLeShootIng
error indicators

Indication Cuff pressure reaches 330mmHg

Cause

Twisted bladder inside cuff

Correction Ensure that bladder lies flat within cuff

Indication "Err" displays, cuff pressure releases, then "0 pressure" displays

Cause

Excessive arm movement or talking during measurement

Correction Turn monitor OFF then ON and measure again

while remaining still

Indication Could not obtain pulse rate

Cause

Cuff was too loosely applied, extremely weak pulse, or arrhythmia

Correction Reapply cuff and measure again; if problem persists after cuff is

correctly positioned, seek medical attention immediately

Batteries are exhausted. Replace with four new AAA alkaline batteries

beverages, before a measurement.

4. Remove any clothing on the upper arm to place the

cuff directly on the skin. Constriction of the upper
arm by clothing or a rolled-up sleeve, as shown at
right, may cause an inaccurate reading; remove any
constricting garments.

5. Sit down in a chair with back support

and place both feet on the floor.

6. Extend the arm from which the blood

pressure will be taken to the front or
side. Rest your hand, palm-up, on a
table as shown at right.

Note: Place the cuff at the same level as your heart to obtain accurate
measurements. Your heart is located slightly to the right of your left
armpit.