Graham Field NEB-U-TYKE JB0112-164 User Manual

Page 11

Advertising
background image

John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 11

Storage

Keep the device and accessories dry. Avoid exposure to
direct sunlight.

TROUBLESHOOTING
If your Neb-u-Tyke RR Train Nebulizer Compressor fails
to function, consult the following Troubleshooting guide.
If the problem persists, consult your Graham-Field autho-
rized distributor.

Problem

Cause and Solution

Device doesn’t
operate

Check if plug is properly fit into a grounded electrical outlet.

When device has been run continuously for more than 30 minutes
just before using, auto-shut-down may have been activated by the
built-in thermal protector; ensure that device is OFF for at least 30
minutes before next usage.

Nebulization is
weak

Check for proper electrical voltage.

Check tubing for blockage or air leakage at connection to Neb-u-Tyke
compressor or nebulizer cup; replace as needed.

Check that the nebulizer cup is properly assembled and not
damaged. If there is any damage, replace as needed.

Check if filter is dirty; replace filter as needed.

Advertising