Graham Field OXYCHECK JB02017 User Manual

Page 2

Advertising
background image

2

SPECIFICATIONS

Display Type

LED (Light Emitting Diode)

SpO

2

Measurement

range

70-99%

Accuracy

80%-99%: ±2%

70%-79%: ±3%

≤69% : no definition

Pulse Rate

Measurement

range

30-235 BPM

Accuracy

30~99 BPM: ±2 BPM
100~235 BPM: ±2%

Pulse Intensity

Bargraph Indicator

Power Requirement Two AAA alkaline Batteries
Power Consumption <40 mA
Low Power Indicator

Battery Life

~ 30 hours of continuous operation

Dimension

(L x W x H)

2.20

~ 2.44 x 1.26 ~ 1.50 x 1.34 ~ 1.50

(56 mm ~ 62 mm x 32 mm ~ 38 mm x 34 mm ~ 38 mm)

Weight

1.59 oz ~ 2.12 oz (0.10 lb ~ 0.13 lb) (45 g ~ 60 g)

including two AAA batteries

Environmental

Requirements

Temperature

Operation

41°F ~ 104°F

(5°C ~ 40°C)

Storage

68°F ~ 131°F

(20°C ~ 55°C)

Humidity

(non-condensing)

Operation

≤80% RH

Storage

≤93% RH

Interference

Resistance Capacity

against Ambient Light

Device works normally when mixed noise produced

by BIO-TEK INDEX Pulse Oximeter tester

DECLARATION

This product’s EMC complies with IEC60601-1-2 standard.

The materials with which the user can come into contact have

no toxicity, no action on tissues, and comply with ISO10993-1,

ISO10993-5 and ISO10993-10.
GUIDANCE AND MANUFACTURER’S DECLARATION:
ELECTROMAGNETIC EMISSIONS FOR ALL EQUIPMENT
AND SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic emission
The Pulse Oximeter is intended for use in the electromagnetic

environment specified below. The customer or the user of the Pulse

Oximeter should assure that it is used in such an environment.

Emission Test Compliance Electromagnetic environment – guidance
RF emission
CISPR 11

Group 1

The Pulse Oximeter uses RF energy only for its

internal function. Therefore, its RF emissions

are very low and are not likely to cause any

interference in nearby electronic equipment.

RF emission
CISPR 11

Class B

The Pulse Oximeter is suitable for use

in all establishments, including domestic

establishments and those directly connected to

the public low-voltage power supply network that

supplies buildings used for domestic purposes.

TROUBLESHOOTING

Problem

Possible reason

Solution

SpO

2

or PR

cannot be

displayed

normally

1. User's finger is incorrectly

inserted

2. User’s Oxyhemoglobin

value is too low to be

measured

1. Reinsert user's finger
2. Attempt several times to

obtain a reading; If sure that

no problem exists, obtain

further clinical examination

SpO

2

or PR

display is

unstable

1. Finger may not be

inserted deeply enough

2. Finger trembling or user

moving

1. Reinsert finger
2. Ask user to remain still

Troubleshooting continued

Problem

Possible reason

Solution

Oximeter

cannot be

powered on

1. Battery power may be

inadequate or batteries

may not be installed

2. Batteries may be installed

incorrectly

3. Oximeter may be damaged

1. Replace batteries
2. Reinstall batteries
3. Contact GF distributor

Indicator

lamps are

suddenly

off

1. Device automatically

powers off when no signal

is detected for longer than

8 seconds

2. Batteries too weak to

power device

1. Normal
2. Replace the batteries

“Error3” or

“Error4”

displays

1. Low power
2. Receiving tube and/or

connector may be

shielded or damaged

3. Mechanical misplace for

receive-emission tube

4. Amp circuit malfunction

1. Replace batteries
2. Contact GF distributor
3. Contact GF distributor
4. Contact GF distributor

“Error7”

displays

1. Low power
2. Emission diode damaged
3. Current control circuit

malfunction

1. Replace batteries
2. Contact GF distributor
3. Contact GF distributor

SYMBOL DEFINITIONS

Symbol

Definition

Symbol

Definition

Type BF applied part

Low power indicator

Attention, consult

accompanying documents

No SpO

2

Alarm

﹪ SpO

2

Oxygen saturation

Power switch

BPM

Heart rate (BPM)

SN

Serial Number

Info: The illustration used in this manual may differ slightly

from the appearance of the actual product.

LIMITED WARRANTY

GF Health Products, Inc. warrants the John Bunn

®

JB0217 Oxy-

Check Fingertip Pulse Oximeter to be free from defects in work-

manship and materials for one year.
The warranty does not apply to damage resulting from failure to

follow the operating instructions, accidents, abuse, alterations or

disassembly by unauthorized individuals.
If a product is deemed to be under warranty, GF Health Products,

Inc. shall provide, at its option, (1) replacement of any defective

part or product or (2) a credit of the original selling price made to

GF Health Products, Inc.’s initial customer. The warranty does not

include any labor charges incurred in replacement part(s) instal-

lation or any associated freight or shipping charges to GF Health

Products, Inc.
The warranties contained herein contain all the representations

and warranties with respect to the subject matter of this document,

and supersede all prior negotiations, agreements and understand-

ings with respect thereto. The recipient of this document hereby

acknowledges and represents that it has not relied on any rep-

resentation, assertion, guarantee, warranty, collateral contract or

other assurance, except those set out in this document.

www.grahamfield.com

Graham-Field and John Bunn are registered trademarks of GF Health Products, Inc.