Connecting the water inlet

,

Water in the machine is not

suitable for drinking.

– The machine must be connected to

the water supply in accordance with
local and national regulations.

– High iron content can leave a rust

film on stainless steel labware and
the machine itself.
When the chloride content in the
domestic water supply is above
100mg/l, the risk of corrosion
damage to stainless steel labware is
greatly increased.

– In certain regions (e.g., mountainous

areas) the water composition may
cause condensate to form, requiring
the use of softened water in the
steam condenser.

– An acceptable water pressure (flow

rate) is 10 - 147 psi. However, the
recommended pressure is
25 - 60 psi.
If the water pressure is below 30 psi,
the fill time will take longer.
If the "Fill/Drain" indicator fault code
"F..E" lights, contact the Miele
Technical Service Department.

– The maximum permissible static

water pressure is 145 psi over
pressure.

– If the water pressure lies outside

these ranges, please contact Miele
Technical Service for advice (the
"Fill/Drain" indicator, may appear in
the display "F E").

– The machine is designed for

connection to cold water (blue
marking) and hot water (red marking)
up to 158°F (70°C) max.
Connect the water intake hoses to
the cold and hot water faucets.

– If there is no hot water supply, the

intake hose with the red marking
must also be connected to the cold
water supply.

– Connect the intake hose for the

steam condenser (without water
safeguard) to the cold water faucet.

– See the next page for how to

connect the DI water line.

– Water valves with

3

/

4

inch thread are

to be provided on site. They should
be easily accessible, because the
water supply will need to be turned
off when not in use.

– The inlet hoses are approx. 5 ½ ft

(1.7 m) long pressure hoses with

3

/

4

inch thread. Do not remove the

water inlet filters in the threads.