Pressure Systems, Inc. Model 9116 User’s Manual

Page 68

www.PressureSystems.com

Example:

●

Send TCP/IP command to Model 9116 module (via its connected socket) that

returns decimal voltage data (of the raw temperature signal) for channels 1, 5, 9,

and

13:


“n11110”


Response contains data for channels 13, 9, 5, and 1 (left to right):


“ 0.53013 0.541698 0.503633 0.000000”


In this example channels 13, 9, and 5 return normal temperature voltage signals

in the range of 0.5 to 0.6 volts. Note that channel 1 returns a value of 0.0 volts,

indicating a possible error in its temperature signal.