SECTION 4—CLEANING AND DISINFECTING

Part No 1130202

29

Stratos™Portable

Cleaning the Nebulizer

ƽ WARNING

To prevent possible risk of infection due to

contamination, follow these instructions:
• Clean the nebulizer before initial operations.
• Clean the nebulizer after each aerosol treatment.
• ALWAYS use fresh cleaning solution.
• DO NOT towel dry the nebulizer components after

cleaning.

NOTE: Your physician and/or an Invacare dealer may specify

certain cleaning procedures. If so, follow their recommendations. If

not, clean according to the following instructions:
1. Remove the nebulizer and mouthpiece from the tubing.
2. Disassemble nebulizer chamber by turning

counterclockwise and separating.

3. Fill two plastic containers or bowls:

A. Washing: Hot water/detergent solution.
B. Soaking: Hot water/vinegar solution (one part vinegar

to three parts water).