Ingersoll-Rand DOSER 45532694 User Manual

Page 10

Advertising
background image


10

45532694_ed2

m) Place the cylinder (3) on the piston, (4). Align it, and press it into the housing.

n) Position all parts on the docking valve according to the drawing. Tighten the four screws (19).

Torque is 10 Nm.

o) Finally, the pneumatic unit function should be tested. Connect the air and check the movement of

the needle.

1.7 Female Docking Valve


See “Doser” Maintenance Manual form 45532660.