Chapter 3

Options

3-2

3.2 Four Sensor Card

The SQM-160 automatically detects the presence of the Four Sensor Card.

Refer to System Connections in Chapter 1 for sensor hookup information.

Refer to Chapter 2, Sensor Selection, to assign Four Sensor Card sensors to a film.

3.3 Analog Output Card

Once the Analog Output Card is installed, the SQM-160 System Parameters must be
set to match the device that will be attached to the Rate or Thickness output.

To setup the Analog Output Card in the System Parameters menu:

1. Press Program to enter program mode.

2. Press Prev to enter the System Parameters Menu

3. Use Next to move through the system parameters until R/T BNDS is displayed.

4. Depress the Control Knob to display RATE MIN.

5. Adjust the Control Knob to the Rate desired for a 0V output.

6. Depress the Control Knob to save the value and display the RATE MAX setting.

7. Adjust the Control Knob to the Rate desired for a 5V output.

8. Repeat steps 4-7 to adjust the Thickness output values.

9. Press Program to exit the System Parameters Menu and return to normal mode.

Refer to System Parameters Setup in Chapter 2 for more information on setting SQM-
160 System Parameters.