Sentry Industries 5000 User Manual

Page 82

Advertising
background image

SMC

sierra monitor corporation

Sentry Instruction Manual - Version 6

SERVICE

Page: 76

7.8 HYDROGEN SULFIDE MODULE (5100-05)

7.8.1 DESCRIPTION

Model 5100-05 Hydrogen Sulfide Sensor Module
includes a sensor assembly and electronic
assembly installed in an explosion proof housing.

The sensor assembly includes a reuseable housing
and disposable electrochemical sensor. The
assembly screws into one hub of the sensor
module enclosure and plugs into the bottom
electronics card via a six pin connector.

Cabling from the controller connects to a three pin
spring loaded terminal strip on the electronics
assembly.

7.8.2 TROUBLE ANALYSIS

Electrical adjustment, or replacement of the sensor
will be necessary under the following conditions:

Controller displays the following error
messages

SENSOR FAILURE

LOW SENSITIVITY

False readings or alarms are received
due to sensor inaccuracy.

Warning: During sensor adjustments the
concentration reading on the controller will be inaccurate
and alarm level concentrations may be displayed. If false
activation of the alarm relays will cause a problem
disconnect the relay wiring prior to adjustment or turn the
module off using the "Change Module" mode.

NOTE

Although all the necessary data can be collected with a

voltmeter at the sensor module, some helpful

information can be displayed on the controller or

printed. See diagnostic codes 0004 and 0008 in

Appendix C.

7.8.3 ADJUSTMENT PROCEDURE

Prior to reading voltages and making adjustments perform
a visual inspection to confirm that there are no physical
problems such as water in the electronics enclosure,
wiring damage or corrosion.

Use Figure 7-9 to locate test points during the following
procedures.

7.8.3.1 SIGNAL ADJUSTMENT

It is unlikely that any electrical adjustment will be
required except when a new sensor is installed.

To make the electrical adjustment connect a voltmeter to

SIGNAL OUT

(pos) and

GND

(ground) and use

SPAN ADJ

potentiometer to adjust the voltage based on the
following:

1 ppm H

2

S = 0.02 VDC.

100 ppm H

2

S = 2.00 VDC.

7.8.3.2 SENSOR REPLACEMENT

The Hydrogen Sulfide sensor should be replaced when it
can no longer be calibrated correctly. Generally this is
every twenty four to thirty months.

The

SENSOR FAILURE

message for Hydrogen Sulfide

sensors does not indicate a specific failure of the sensor
but indicates that the sensor is not correctly connected to
the electronics. If this message appears check that the
sensor harness is correctly installed to the connector on
the bottom electronics board.

The gas sensor which is located inside the sensor
assembly housing can be replaced without replacement of
the housing. To replace the sensor:

Figure 7-9

Cover Plate - Hydrogen Sulfide Module

Advertising