6 Twin Guard Series

™

-86°C Ultra-Low Freezers

www.twinguardseries.com

Twin Guard

™

Dual°Cool Refrigeration System

Dual, independent refrigeration systems operate in Normal or energy-saving EcoMode™ which is selected by the user and engaged through the
LCD controller. Normal Mode is used for the strictest of GMP applications. EcoMode™ is applicable to 90-95% of applications.

Twin Guard

™

Dual°Cool Refrigeration System

Status

Function

Performance

System A and B ON

Maximum Pull-Down and Recovery Capcity

Establishes highly uniform -86°C storage

temperature; maximizes recovery following

door openings and heat load additions

B

-86˚C

A

System A and B ON, Cycling On/Off

EcoMode™

Maintains better energy management at

high or low ambient temperatures as well as

excellent top-to-bottom uniformity.

B

-86˚C

A

System A and B ON, Cycling On/Off

Normal Mode

Maintains excellent top-to-bottom uniformity.

Maintains most repeatable, cycling wave form

for the strictest of GMP applications.

B

-86˚C

A

System A ON

Back-Up for System B

Maintains minimum -65°C reserve

temperature

B

-65˚C

A

System B ON

Back-Up for System A

Maintains minimum -65°C reserve

temperature

B

-65˚C

A

The SANYO Twin Guard Series

™

Freezer can be set for Normal or

EcoMode

™

operation. EcoMode

™

establishes an overlapping cycle to

significantly reduce energy consumption.

2 compressors are turned ON/OFF at same time

2 compressors and fan motor for each are independently controlled by SV and read AT.

Chamber

Comp A
Comp B

temp

T12 on

T12 off

Normal

Normal

Chamber

Comp A
Comp B

temp

T12 on

T12 off

T2 on

T2 off

Chamber

Comp A
Comp B

temp

T on

T1 off

T12 on

T2 off

T1 2 on

ECO mode

ECO mode

ECO mode operation image

Pull down

Temp.

Comp A
Comp B

Stable temperature

NORMAL mode operation image

Pull down

Temp.

Comp A
Comp B

Stable temperature

This chart shows the repeatable cycling wave form of normal mode vs. eco mode.
Normal mode is for strict GMP applications.

EcoMode

™

vs. Normal Mode