Appendix D:
Program Examples

All of the program examples which follow do the
same thing, only the computer, language, or
interface is changed. The programs read the
Channel 1 and 2 Outputs and write the results to
the computer screen. In addition, the X6 analog
output port is ramped from 0 to 10V.

Program Example 1:
IBM PC, Basic, via RS232

In this example, the IBM PC's ASYNC port (known
as COM1: or AUX: to DOS users) will be used to
communicate with the SR510. Only two wires
between the IBM PC's ASYNC port and the
SR510 are needed (pins #2 & #3 of the RS232),
but pins 5,6,8 and 20 should be connected
together on the connector at the IBM end.

10

′

EXAMPLE PROGRAM TO READ THE SR510 OUTPUT AND RAMP THE X6 ANALOG OUTPUT

20

′

USING IBM PC BASICA AND THE COM1: RS232 PORT.

30

′

40

′

50

′

ON THE REAR PANEL OF THE SR510, SET SWITCH #1 OF SW2 DOWN

60

′

AND ALL OTHER SWITCHES IN SW2 UP. (9600 BAUD, NO PARITY)

70

′

80 OPEN

″COM1

:9600,N,8,2,CS,DS,CD

″ AS #1

90

′

SET UP COM1: PORT TO 9600 BAUD, NO PARITY, 8 DATA BITS, 2 STOP BITS,

100

′

IGNORE CTS (CLEAR TO SEND), DSR (DATA SET READY),

110

′

AND CD (CARRIER DETECT).

120

′

130 PRINT #1,

″

″

′CLEAR UART BY

SENDING SPACES

140 PRINT #1,

″Z″

′RESET SR510

150 FOR I = 1 TO 200: NEXT I

′WAIT FOR RESET TO FINISH

160

′

170 X = 0

′INIT X6 OUTPUT TO ZERO

180

′

190 PRINT #1,

″Q″ ′READ OUTPUT

200 INPUT #1,V1

′INTO V1

210

′

220 PRINT

″OUTPUT = ″

;V1

230

′

240 X =X + .0025

′INCREMENT X6 OUTPUT BY 2.5 MV

250 IF X > 10 THEN X = 0

′RESET X6 RAMP

260 PRINT #1, USING

″X6, ##

.###

″;X

′SET X6 OUTPUT VOLTAGE

270

′

280 GOTO 190

′LOOP FOREVER