DAILY INSPECTION INSTRUCTIONS

3.4

Figure 3.3 Sealed source in transit mode

Sealed source

Source assembly

Shield

Exposure device

Locking

mechanism

Outlet port

Source

guide tubes

Control cable

Remote

control crank

Second, during termination of a radiographic exposure, the remote controls are used to

retract the source back into the exposure device. When the source assembly is returned to

the fully shielded position within the exposure device, the locking mechanism automatically

secures the source assembly, preventing the source assembly from being inadvertently

exposed from the fully shielded position. The automatic securing mechanism allows

movement of the exposure device with the remote controls attached at the job site while

preventing inadvertent exposure of the source assembly.
For the third feature, the controls cannot be removed from the exposure device locking

mechanism unless the source assembly has been returned to the fully shielded position

within the exposure device.
The complete radiography system consists of the radiographic exposure device containing

the source assembly, the remote controls, source guide tubes and collimators (beam limiters)

whenever possible.
The use of multiple source guide tubes must be a combined length shorter than the length

of the remote controls, e.g. if the remote controls are 25ft (7.6m) long, only 3 source guide

tubes can be utilized for a combined length of 21ft (6.4m). This consideration ensures the

source assembly will be fully projected to the source stop (exposure head).
The source assembly must never be projected until the equipment is properly connected and

all personnel have left the restricted area.
Transit mode
Rotating the handle of the remote control crank in the EXPOSE direction (counter-clockwise)

moves the source assembly out of the shielded position in the exposure device and into the

source guide tubes (projection sheaths). (See figure 3.3)
The source assembly is under positive mechanical control using the remote control crank.

The remote controls that are fitted with an odometer to provide an indication of the sealed

source’s position. The odometer reading should never be a substitute for the requirement

and proper use of a radiation survey meter.