SRS Labs SR785 User Manual

Page 12

Advertising
background image

Flatness

<0.25 dB pk-pk (typical),
<1.0 dB pk-pk (max), 5000 rms
averages

Pink Noise Source

Bandwidth

DC to 102.4 kHz

Flatness

<2.0 dB pk-pk, 20 Hz - 20 kHz
(measured using averaged 1/3 octave
analysis)

Chirp Source

Time Record

Continuous or Burst

Output

Sine sweep across the FFT span

Flatness

±0.25 dB, 1.0 Vpk

Swept Sine Source

Auto Functions

Source level, input range and
frequency resolution

Dynamic Range

145 dB

Arbitrary Source

Amplitude Range

±5 V

Record Length

2 Msamples (playback from arbitrary
waveform memory or capture buffer).
Variable output sample rate.

General

Monitor

Monochrome CRT, 800H by 600V
resolution, 8.2 inch diagonal display

Interfaces

IEEE-488, RS-232 and Printer
interfaces standard.
All instrument functions can be
controlled through the IEEE–488 and
RS-232 interfaces. A PC (XT)
keyboard input is provided for
additional flexibility.

Hardcopy

Print to dot matrix and PCL
compatible printers. Plot to HP-GL or
Postscript plotters. Print/Plot to
RS-232 or IEEE–488 interfaces or to
disk file. Additional file formats
include GIF, PCX and EPS.

Disk

3.5" DOS compatible format, 1.44
Mbytes capacity. Storage of displays,
setups and hardcopy data.

Preamp Power

Power connector for SRS
preamplifiers.

Power

70 Watts, 100/120/220/240 VAC,
50/60 Hz

Dimensions

17"W x 8.25"H x 24"D

Weight

56 lbs.

Warranty

One year parts and labor on materials
and workmanship.

Transient Capture

Mode

Continuous real-time data recording to
memory.

Maximum Rate

262,144 samples/sec for both inputs

Max. Capture Length

2 Msamples (single input)
8 Msamples with optional memory

Octave Analysis

Standards

Conforms to ANSI standard
S1.11-1986, Order 3, Type 1-D.

Frequency Range

Band centers:
Single Channel
1/1 Octave

0.125 Hz - 32 kHz

1/3 Octaves

0.100 Hz - 40 kHz

1/12 Octaves

0.091 Hz - 12.34 kHz

Two Channels
1/1 Octave

0.125 Hz - 16 kHz

1/3 Octaves

0.100 Hz - 20 kHz

1/12 Octaves

0.091 Hz - 6.17 kHz

Accuracy

<0.2 dB (one second stable average,
single tone at band center)

Dynamic Range

80 dB (1/3 octave, two second stable
average) per ANSI S1.11-1986

Sound Level

Impulse, Peak, Fast, Slow and Leq per
IEC 651-1979 Type 0

Order Tracking

Resolution

Up to 400 lines

Amplitude Accuracy

±1 dB (typ.)

Displays

Order map (mag and phase), order
track (mag and phase), orbit

Curve Fit and Synthesis

Type

20-pole/20-zero non-iterative rational
fraction fit.

Output Format

Pole-zero, polynomial, or pole-residue

Source Output

Amplitude Range

0.1 mVpk to 5 Vpk

Amplitude Resolution

0.1 mVpk

Offset Adjust

±5 V

Output Impedance

<5

, ±100 mA peak output current

Sine-Source

Amplitude Accuracy

±1% of setting, 0 Hz to 102.4 kHz
0.1 Vpk to 5.0 Vpk, high impedance
load

Harmonics, SubHarm.

0.1 Vpk to 5 Vpk

and Spurious Signals

<-80 dBc (fundamental <30 kHz)
<-75 dBc (fundamental <102 kHz)

Two Tone Source

Amplitude Accuracy

±1% of setting, 0 Hz to 102.4 kHz,
0.1 Vpk to 5 Vpk, high Z load

Harmonics, SubHarm.

<-80 dBc, 0.1 Vpk to 2.5 Vpk

White Noise Source

Time Record

Continuous or Burst

Bandwidth

DC to 102.4 kHz or limited to span

Advertising