Konica Minolta ILLUMINANCE METER LUMINANCE METER CHROMA METER T-10A /T-10MA/ T-10WSA/T-10WLA User Manual

Page 3

Advertising
background image

4

5

PWM is the abbreviation of Pulse Width Modulation, and refers to

the method of controlling signal intensity by controlling the ratio

between the ON period and OFF period of a pulse signal.

A pulse signal is a signal which repeatedly alternates between ON

and OFF, and the percentage of ON period during a single cycle is

referred to as the "duty cycle".

PWM-controlled lighting is a method for controlling the brightness

of a lamp by controlling the duty cycle (lit time) of light from a

pulse-emission source. As the lit time becomes longer, the light

becomes brighter, and conversely, as the lit time becomes shorter

the light becomes darker.

Period when LED is lit

Convenient, easy-to-use Excel

®

add-in software

Reads measurement data from T-10A series Illuminance

Meters directly into Excel

®

. Further processing of data can

then be performed easily using the various functions of Excel

®

.

Main specifications of Data Management

Software T-S10w

Type

Add-in for Excel

®

(Excel

®

is required to use this add-in.)

Operating

environment

One of the following environments with Excel

®

installed:

* Languages in parenthesis ( ) are the OS language.

Windows

®

XP + Excel

®

2003 (English, Japanese, or Simplified

Chinese)

Windows

®

7 + Excel

®

2010 (English, Japanese, or Simplified

Chinese)

* For details on system requirements for above versions of

Windows

®

and/or Excel

®

, refer to their respective specifications.

* Not compatible with 64-bit versions of office 2010.

Compatible

instruments

T-10A, T-10MA, T-10W

S

A, T-10W

L

A, T-10, T-10M,

T-10W

S

, T-10W

L

Data Management Software T-S10w (Optional accessory)

About PWM-controlled lighting

Relative Spectral Response

Cosine Correction Characteristics

Mini plug

Mini Cap

T-A14

(for T-10MA)

<

Standard accessories

>

<

Optional accessories

>

AA-size batteries

2 pcs.

Strap

Case

T-A10

Cap (with Strap)

T-A13

(for T-10A)

Personal computer

(Commercial product)

RS-232C printer

(Commercial product)

Data Management software

T-S10w

(Including

T-A15)

Data management using computer

Data recording using printer

USB Cable

T-A15

Printer Cable

T-A12

AC Adapter

External power

T-10A

Receptor

Head

Adapter Unit for

Main Body

T-A20

Adapter Unit for

Receptor Head

T-A21

With LAN category 5 cable; 1m

For multi-point and cable extension measurement

Example of

multi-point System

Example of cable

extension

LAN (10 BASE-T)

category 5 straight cable

(commercially

available)

Multi-point measurement requires use of

optional AC adapter.

Hard Case

CL-A10

Hard Case

CL-A10

Additional receptors (sold separately; Product

includes 1 receptor according to model purchased)

Custom models

T-10A

T-10MA

T-10W

S

A

T-10W

L

A

Custom order

T-10MA Receptor Head

T-10W

S

A Receptor Head

T-10W

L

A Receptor Head

Main Specifications of T-10A

Model

Illuminance Meter T-10A

(Standard receptor head)

Illuminance Meter T-10MA

(Mini receptor head)

Illuminance Meter T-10W

S

A

(Waterproof mini receptor head)

Illuminance Meter T-10W

L

A

(Waterproof mini receptor head)

Type

Multi-function digital illuminance meter with detachable receptor head (Multi-point measurements of 2 to 30 points is possible)

Illuminance meter class

Conforms to requirements for Class AA of JIS C 1609-1: 2006 "Illuminance

meters Part 1: General measuring instruments" Conforms to DIN 5032 Part

7 Class B

Conforms to requirements for special Illuminance meters of JIS C 1609-1:

2006 *

1

Receptor

Silicon photocell

Relative spectral response Within 6% (f

1

´) of the CIE spectral luminous efficiency V (λ)

Cosine response (f

2

)

Within 3%

Within 10%

Measuring range

Auto range (5 manual ranges at the time of analog output)

Measuring function

Illuminance (lx). illuminance difference (lx). illuminance ratio (%). integrated illuminance (lx·h).

integration time (h). average illuminance (lx).

Measuring

range

Illuminance 0.01 to 299,900 lx; 0.001 to 29,990 fcd

1.00 to 299,900 lx; 0.1 to 29.990 fcd *

2

Integrated

illuminance

0.01 to 999,900 x 10

3

lx·h 0.001 to 99,990 x 10

3

fcd·h / 0.001 to 9999 h

User calibration function

CCF (Color Correction Factor) setting function: Measurement value x 0.500 to 2.000

Linearity

±2% ±1 digit of displayed value

Temperature/

humidity drift

Within ±3%

Computer interface

USB

Printer output

RS-232C

Analog output

1 mV/digit, 3 V at maximum reading; Output impedance: 10 KΩ; 90% response time: 28 ms

Display

3 or 4 Significant-digit LCD with backlight illumination (Automatic illumination)

Power source

2 AA-size batteries / AC adapter AC-A308 (optional; for 1 to 10 receptors) or

AC adapter AC-A311 (optional; for 1 to 30 receptors)

Battery life

72 hours or longer (when alkaline batteries are used) in continuous measurement

Operating temperature

/humidity range

-10 to 40°C, relative humidity 85% or less

(at 35°C) with no condensation

5 to 40°C, relative humidity of 85% or less

(at 35°C) with no condensation

Storage temperature /

humidity range

-20 to 55°C, relative humidity 85% or less

(at 35°C) with no condensation

0 to 55°C, relative humidity of 85% or less

(at 35°C) with no condensation

Dimensions

69 x 174 x 35 mm

Main body: 69 x 161.5 x 30 mm

Receptor: Ø16.5 x 13.8 mm

Cord length

1 m

5 m

10 m

Weight (without battery) 200 g (7.0 oz.)

205 g

260 g (Receptor head only: 120 g)

340 g (Receptor head only: 200 g)

*1 Conforms to requirements for Class AA of JIS C 1609-1: 2006 for all items except cosine response (f

2

).

*2 Although measurements below 1.00 lx are possible, they may not be stable due to the effects of electrical noise.

<Notes regarding mini receptors and waterproof mini receptors>

*Do not touch the cable during measurements. Doing so may result in unstable measurement values.

*Secure the cable during measurements. Failure to do so may result in unstable measurement values.

Center of

tripod

socket

Center of

receptor window

Center of

receptor

window

7.5

T-10MA 1 m

T-10W

S

A 5 m

T-10W

L

A 10 m

25

35

174

117.5

161.5

69

45.2

T-10A

T-10MA/ T-10W

S

A/ T-10W

L

A

30

16.5

13.613.9

14

Reference

plane

Reference

plane

12.7

3

6.5

Center of

tripod

socket

Center of

receptor window

Center of

receptor

window

7.5

T-10MA 1 m

T-10W

S

A 5 m

T-10W

L

A 10 m

25

35

174

117.5

161.5

69

45.2

T-10A

T-10MA/ T-10W

S

A/ T-10W

L

A

30

16.5

13.613.9

14

Reference

plane

Reference

plane

12.7

3

6.5

Dimensions

(Units: mm)

System diagram

Illuminance Meter T-10A series

Illuminance Meter T-10A series

Ideally, the relative spectral responsivity of the illuminance

meter should match V (λ) of the human eye for photopic vision.

As shown in the graph above, the relative spectral responsivity

of Konica Minolta Illuminance Meters T-10A/10MA is within

6% (f1´) of the CIE spectral luminous efficiency V (λ).

CIE ; Commission Internationale de I´Eclairage
f1´(CIE symbol) ; The degree to which the relative spectral

responsivity matches V (λ) is characterized by means of the error f1´.

Since the brightness at the measurement plane is proportional

to the cosine of the angle at which the light is incident, the

response of the receptor must also be proportional to the

cosine of the incidence angle.

For Konica Minolta Illuminance Meters T-10A/10MA, the

cosine response f2 is within 3%.

The graph above shows the cosine correction characteristics

of Konica Minolta Illuminance Meters T-10A/10MA.

Relative sensitivity (%)

Wavelength (nm)

Konica Minolta Illuminance

Meters T-10A/T-10MA

The spectral

luminous efficiency

(λ)

400

450

500

550

600

650

700

750

0

10

20

30

40

50

60

70

80

90

100

60

60

40

40

20

20

0

100

%

80

%

60

%

40

%

20

%

80

80

Konica Minolta Illuminance

Meters T-10A/10MA

Ideal curve

Advertising